Electromagnetic Environment-guidance
Portable and mobile RF communication equipment should be used no
closer to any part of the 18708, including cables, than the recommend-
ed separation distance calculated from the equation applicable to the
frequency of the transmitter. Recommended Separation Distance:
Immunity Test
IEC60601 Test level
Compliance level
Radiated RF
ICE 61000-4-3
3V/m 80MHz to 2.5GHz
3V/m
√P
80MHz to 800 MHz
E
1
d =
[
3.5
]
√P
800MHz to 2.5 GHz
E
1
d =
[
7
]
50