4-12
Communication Interfaces
CESAR 1312 (E)
Dressler
®
HF-Technik GmbH
CESAR generator and the controller is 10 meters (33 feet). Minimize
interference from adjacent electrical equipment by terminating the EMI shield
in the cable to the metal shells of the cable’s connectors. Additionally, you
must tie the chassis of the CESAR generator to a local earth ground through
an adequately sized copper grounding strap.
Note:
Grounding the
User Port
at the CESAR generator reduces noise
interference. To avoid ground loop problems, you should typically
ground only one end of the
User Port
cable.
ACTIVATING THE 25-PIN USER PORT
To activate the 25-pin User Port control
1. Press the
Program
soft key.
2. Press
Next
until you see
Device Configuration
on the display.
3. Press
Change
.
4. Press
OK
until you see
Control by
on the right-hand side of the screen.
5. Press
User Port
.
6. Press
Execute
.
Summary of Contents for CESAR 1312
Page 1: ...U s e r M a n u a l s e r v i c e d r e s s l e r c o m CESAR Generator Model 1312 ...
Page 2: ......
Page 3: ...User Manual CESAR Generator Model 1312 CESAR 1312 E ...
Page 6: ...iv CESAR 1312 E Dressler HF Technik GmbH ...
Page 14: ...xii List of Tables CESAR 1312 E Dressler HF Technik GmbH ...
Page 24: ...1 10 Safety and Product Compliance Guidelines CESAR 1312 E Dressler HF Technik GmbH ...
Page 34: ...3 8 Specifications CESAR 1312 E Dressler HF Technik GmbH ...
Page 170: ...5 50 Installation Setup and Operation CESAR 1312 E Dressler HF Technik GmbH ...
Page 188: ...6 18 Troubleshooting and Customer Support CESAR 1312 E Dressler HF Technik GmbH ...