5-34
Installation, Setup, and Operation
HPG 13150 400 V (A)
Dressler
®
HF-Technik GmbH
Prerelease version
If your HPG generator has an RS232 port, the RS-232 protocol is factory-
set to
AE Bus
, and most HPG generators will use the default AE Bus
protocol (see
“AE Bus Protocol” on page 4-23
). However, a previous
protocol (
Dressler
) is available for compatibility.
Note:
To determine if you have an RS-232 port, see
Model 13150 400 V Options” on page 1-2
To Change the Default Device Configuration Settings Using the HPG Generator
Front Panel:
1. Press the
Program
soft key.
2. Press
Next
until you see
Device Configuration
on the display.
3. Press
Change
.
4. Select the desired
Displ. Contrast
value using the
RF POWER
rotary dial
knob.
5. Press
Ok
to confirm.
6. Press the desired
Analog I/O Max
level using the
RF POWER
rotary dial
knob.
4
You may select values between 2 V and 10 V.
7. Press
Ok
to confirm.
8. Select the desired
DC Bias Scaling
using the
RF POWER
rotary dial
knob.
4
You may select values between 0 V and 4000 V using a combination of
the
Increm Auto
key and the
RF POWER
rotary dial.
9. Press
Ok
to confirm.
10. Press the desired
Control by
setting:
4
Front Panel
defaults to front panel control.
4
User Port
defaults to
User Port
control.
11. If your HPG generator has an RS-232 port, select the desired
RS232
Baud Rate
value using the
RF POWER
rotary dial knob.
12. Press
Ok
to confirm.
13. If your HPG generator has an RS-232 port, select the desired
RS232
Protocol
.
4
AE BUS
is the standard protocol for HPG generators.
4
Dressler
is available for compatibility only.
14. Press
Ok
to confirm.
15. Press the
Exit
soft key.
16. Press the
Execute
soft key.
Summary of Contents for 13150 400 V
Page 2: ......
Page 3: ...User Manual Prerelease version HPG Generator Model 13150 400 V HPG 13150 400 V A ...
Page 6: ...iv HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Page 12: ...x List of Figures HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Page 14: ...xii List of Tables HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...
Page 104: ...4 66 Communication Interfaces HPG 13150 400 V A Dressler HF Technik GmbH Prerelease version ...