2. The clear water inside the water tank pot should be frequently replaced, with the pot
cleaned.
3. Electrical part must be resistant against water inflow and moisture.

VII. Failure analysis and exclusion methods

Phenomena

Check (causes)

Handling method

No power supply

indicator

light

does not work

1. Check whether the power plug is

loose

2. Check whether the fuse is

disconnected

1. Insert the power plug tight
2. Replace the fuse

Principal machine
does not rotate

Ditto

Ditto

Water tank can not
be ascended or
descended

1. Check ditto

refers to XD-5000

2.

Check whether the hand

wheel screw is fastened

1. Handle ditto

refer to XD-5000

2. Screw the hanging screw tightly

with hands

Water tank can not
be heated

1 Check the 15A fuse
2. Check whether the heating plug is
loose

1. Replace the 15A fuse
2. Insert the heating plug tightly

Insufficient vacuum

1. Check whether the seal ring is
worn
2.

Check the performance of

vacuum pump, and whether each
connector is tightened (block the air
outlet of vacuum pump to observe
the indicators on vacuum table)
3.

Check

whether

the

glass

appliances are defected or cracked for
coldness
4. Check whether the interface of
glass appliance is coated withvacuum
grease

1. Replace the seal ring
2. Screw the connectors tightly or
replace the vacuum pump
3.Replace

the

damaged

appliances
4. Recoat with vacuum grease

In case of troubleshooting according to above methods, please fax the failure phenomenon

to us at the number of company , instead of operating by yourselves.

VIII. Storage and transportation

Summary of Contents for ZNHW-500ML

Page 1: ... and operation Drawell International Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd PuDong New Area Shanghai China Tel 0086 21 50966080 Fax 0086 21 50966080 Web www drawell com cn Email sales01 drawell com cn ...

Page 2: ...Rotary Evaporator Specification Thank you for choosing XTRACTION SERVICES XD 5000 rotary evaporator our mission make your research more accurate and efficient ...

Page 3: ...e of the device and reduce its performance Therefore please be sure to carefully read the manual before using The manual will be submitted to end users Please keep this manual to prepare for inspection when required The manual is used for only reference the specific shape design is subject to the real object I Usage XD 5000 production type rotary evaporator is a kind of device with the operation p...

Page 4: ...uninstallation 5 Thermostat adopts the touchable smart digital display which can accurately set the heating temperature 6 Frequency control is available for the adjustment of rotation speed with stable and reliable digital display 7 Water bath kettle can be moved up and down The XD 5000 type kettle adopts the method of single hand wheel lift and XD 5000 adopts two ways namely through power and han...

Page 5: ...hard objects 3 Install the four way valve connect the valve to the principal machine encase the forcing screw into four way flange mouth encase the nylon collar align the internal jointing washer and screw the forcing screw tightly 4 Install the condensation pipe insert it into the four way ball milling position and keep vertical screw the fixed packet ring with the slotted screwdriver tightly and...

Page 6: ... of vacuum degree 11 Only after being injected with water can the water bath be powered and heated Operation a Wiring and electricity charging must be done by professional electricians and ground wire terminal must be grounded Power utility power 220v 50Hz b Press the power switch to the state of ON and turn the speed control knob when the principal machine will begin to rotate c After injecting w...

Page 7: ...tly with hands Water tank can not be heated 1 Check the 15A fuse 2 Check whether the heating plug is loose 1 Replace the 15A fuse 2 Insert the heating plug tightly Insufficient vacuum 1 Check whether the seal ring is worn 2 Check the performance of vacuum pump and whether each connector is tightened block the air outlet of vacuum pump to observe the indicators on vacuum table 3 Check whether the g...

Page 8: ...he warranty period for this product is one and half a years commencing from the leaving the factory after being packed well or one year after users purchase Under the premise of using it according to the norms for storage and usage if the product can not work normally we will provide the service of repair or replacement free of charge Glass pieces are beyond the scope of warranty In case of improv...

Page 9: ...tube 1 set 8 Stainless steel ball milling mouth clip 35 2 9 Stainless steel rod bracket 1 10 German cross clip 1 11 Aluminum cross clip Black 1 12 Fixed packet ring 1 13 Bottle supporting rack 1 14 Fuse 20A 2 each 15 Cross screwdriver 1 16 Flask lock wrench 1 17 Certificate warranty card 1 set Product certification Product name rotary evaporator Product model XD 5000 Production date Product number...

Page 10: ...Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd PuDong New Area Shanghai China Tel 0086 21 50966080 Fax 0086 21 50966080 Web www drawell com cn Email sales01 drawell com cn ...

Reviews: