8.8.8.8.8.8.

Battery Voltage U=
Maximum capacity
------------

Finally the balance stays in a standby state of showing 0g, 0.0g in the display

Notice

: the lasting time of --------- will be decided by the stability of the loadcell, thus, the

balance must not be located in an unstable surface or in the wind blow. If O flashes in the
display, it means the balance is not in a stable condition, calibration and counting operations
is not allowed.

CALIBRATION

The purpose for this operation is to calibrate the balance so as to achieve the best
performance in case there is obvious tolerance error in weighing or the balance is
located in different gravity due to different latitude.

It is suggested to warm up the balance over half an hour before calibration.

Remove all loads from the platter, press function key

TARE

to clear the readings to

0g ,0.0g.
Press and hold function key

CAL

for 3 seconds, release function key

CAL

when

“---CAL---“ is shown in the display, a figure of standard weight will flash in the display,
put a standard weight on the platter accordingly, standby state “------“ will be shown in
the display for a few seconds before the standard weight figure is shown in the
display, remove the weight, “------“ standby state will remain for a few seconds before
the balance enter stable weighing mode, showing 0g ,0.0g.
It is recommended to make the operations twice to

achieve the best calibration

result.

WEIGHING

When the balance is warmed up and calibrated, 0g,0.0g is shown in the display

indicating weighing mode, put the object on the platter, when stable, the weight of the object
is shown in the display.

TARING

When a loader is put over the platter, its weight is shown in the display, press function key

TARE

, 0g, 0.0g will be shown in the display, indicating the weight of the loader is deducted,

put the object into the loader, when stable, the figure shown in the display is the weight of
the object.

PIECE COUNTING

Remove all loads from the platter, press function key

TARE

to clear the readings in the

display, press function key

COU

to enter counting mode, figure

10(default

sampling

quantity)

will

be

shown

in

the

display,

weighing

unit