XI. Maintenance

Body cleaning:

please use non-corrosive neutral cleaning agent to clean the inside and outside

surfaces of the ultra low temperature freezer and then use dry cloth to wipe it up.

Notes: it is strictly prohibited to use water to directly wash the inside and outside surfaces of the freezer
so as not to affect the insulating property of electrical equipment. Boiling water, cleaner, acid, alkali,
gasoline, alcohol, benzene, corrosive cleaning agent and scrub-bush are forbidden in cleaning.

Clean dusts on the condenser frequently by banister brush or vacuum cleaner to maintain good
condensing effect.

Regularly wipe a little talcum powder on the door seal to extend the service life of the sealing strip.

If the ultra low temperature freezer has been running for a long time, frost may be easily accumulated on
the door seal, the opening part of the freezer and the side wall inside the freezer. If the frost layer is too
thick, the sealing property and the refrigerating property may be affected. Therefore, please defrost
regularly and wipe it clean with dry cloth.

The water shall not be dumped on the casing of the medical low temperature storage case or kept
indoors,Otherwise, it may damage the electrical insulation and lead to a fault.

Please use non-corrosive neutral cleaning agent to clean the inside and outside surfaces of the freezer
and then use dry cloth to wipe it up.

Do not allow children to use medical freezer to playing games

Not to put heavy objects on the refrigerator door or medical cap, to avoid the door or the top compression
deformation.

Machine halt

: when stop using the freezer, unplug the power plug, clean it according to the above

method, cover it with a plastic bag after natural drying and put it on a ventilated and clean place.

XII. Non-fault phenomenon

Sound of running water can be heard in the ultra low temperature freezer: this is the sound of
refrigerating fluid running in the pipeline.

Compressor is hot: when the compressor is running, the surface temperature can reach to 70

℃~

90

.

You may feel a little hot on the two sides of the body (close to the door body) when the freezer is running.
That is because the leak proof tube equipped in the freezer.

There is condensation on the door seal: in rainy and moist season or under the environment with higher
relative humidity, there may be condensation on the outer surface of the freezer and the door seal. Then
rub dry.

Summary of Contents for MDF-40V50

Page 1: ...and operation Drawell International Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd PuDong New Area Shanghai China Tel 0086 21 50966080 Fax 0086 21...

Page 2: ...0 OPERATION MANUAL MEDICAL FREEZER MDF 40V50 VERSION 2017 4 MDF 40V50...

Page 3: ...s goggles etc 2 Always adopt good health habits and operate strictly according to the product instruction 3 Everyone is obliged to be in charge of one own safety 4 Not Allowed direct contact with pati...

Page 4: ...ination 7 8 1 2 How to see the set point 7 8 1 3 How to change the set point 7 8 1 4 How to start a manual defrost only xr02cx 8 8 1 5 How to change parameter value 8 8 2 parameters 8 8 3 Digital inpu...

Page 5: ...means the place is dangerous to human bodies or the use pattern is harmful to products or alarm users should pay special attention to this It is strictly prohibited to unpack the outer packing case fo...

Page 6: ...ct in the rain or under the environment with the relative humidity of larger than 85 Otherwise faults such as electric leakage etc may be caused With continuous operation system the ultra low temperat...

Page 7: ...Checking standard accessories and data check and compare the inside accessories and data with the packing list please see related pages of the directions Please place the ultra low temperature freeze...

Page 8: ...zer decreases obviously which indicates the refrigerating system is working normally Before storing articles please keep the freezer running in empty state When the freezer is running stably it is bes...

Page 9: ...mode To return to room temperature display LED State Signification On Compressor enabled Flashing Anti short cycle delay enabled AC parameter On Defrost in progress Flashing Dripping in progress On M...

Page 10: ...OINT SET 99 C SET 99 F Set the maximum value for set point ot First probe calibration 9 9 9 9 C 17 17 F allows to adjust possible offset of the first probe od Outputs activation delay at start up 0 99...

Page 11: ...ing dA Exclusion of temperature alarm at startup 0 99 min time interval between the detection of the temperature alarm condition after instrument power on and alarm signalling Digital iP Digital input...

Page 12: ...input is de activated Serious alarm iF bA When the digital input is activated the unit will wait for di delay before signalling the CA alarm message The relay outputs are switched OFF The alarm will s...

Page 13: ...s before replacing the probe Temperature alarms HA and LA automatically stop as soon as the temperature returns to normal values Alarms EA and CA with iF bL recover as soon as the digital input is dis...

Page 14: ...he ultra low temperature freezer is a device used for maintaining low temperature rather than a quick freezing device If you want to store high capacity high water bearing material pre cooling them in...

Page 15: ...d on the casing of the medical low temperature storage case or kept indoors Otherwise it may damage the electrical insulation and lead to a fault Please use non corrosive neutral cleaning agent to cle...

Page 16: ...enance Refrigerant leeks or pipeline is blocked it is necessary to contact the nearby maintenance department for overhaul The condensate fan in the machine room is broken it is necessary to contact th...

Page 17: ...echnology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd PuDong New Area Shanghai China Tel 0086 21 50966080 Fax 0086 21 50966080 Web www drawell com cn Email...

Reviews: