Drawell DW-FD Sereries User Manual Download Page 7

Drawell Scientific

Chapter 2 Technical Specification

I. Technical Specification

Model

FD36R/FD36A FD50R/FD50A FD65R/FD65

A

FD80R/FD80A FD100R/FD100

A

Capacity

36(L)

50(L)

65(L)

80(L)

100(L)

External dimension(mm)

580x490x1080 580x490x1080 730x625x955 730x625x1100 730x625x1200

Size of sterilizer chamber

mm

φ 325x460

φ325x670

φ400x564

φ400x709

φ400x809

Power supply

220V 50/60HZ

Rated power

2300W

2600W

2900W

4600W

4600W

Range of pressure gauge 0-0.4Mpa

Working environment

5-40°C relative humidity 10-85%

Rated working pressure

0.25Mpa

Designed pressure

0.3Mpa

Sterilization temperature

105-135℃

Sterilization timer

1-300min

Drying timer

1-300min ①

Operation mode

Solid Mode、Solid& Drying Mode ②、Liquid Mode

safety equipment

Safety valve, over temperature and temperature abnormal protection
system,over pressure protectionsystem,electric leakage protection equipment,
chamber cover checking system,self- induction interlocking device

Accessories

Stainless steel baskets,waterplate,waste water tank

①② R series have drying mode for solid,drying time can be pre-setted.

Summary of Contents for DW-FD Sereries

Page 1: ...Drawell Scientific USER S MANUAL DW FD Series FD Series Automatic Autoclave...

Page 2: ...oading articles to be sterilized 08 5 Closing the sterilizer chamber cover 09 6 Selecting sterilizing program 09 7 Starting sterilization 10 8 Opening the cover Program finished 10 9 Taking the steril...

Page 3: ...is OMpa before opening the chamber cover Never attempt to open the chamber cover and drainage valve or draw out the steam trapping bottle when the pressure inside the sterilizer chamber is over OMpa...

Page 4: ...peration to allow the instrument to cool down so that it will be able to produce adequate saturated steam In case of abnormal conditions such as noise odor and smog please turn off the power immediate...

Page 5: ...Drawell Scientific Chapter 1 External Features Controller I External Features...

Page 6: ...Drawell Scientific...

Page 7: ...sure gauge 0 0 4Mpa Working environment 5 40 C relative humidity 10 85 Rated working pressure 0 25Mpa Designed pressure 0 3Mpa Sterilization temperature 105 135 Sterilization timer 1 300min Drying tim...

Page 8: ...ake out the water plate remove the packaging place it in the bottom of the chamber in a horizontal position Connectioninstrument must be grounded reliably if the power socket does not 1 The instrument...

Page 9: ...to avoid affecting normal work of sterilization Caution If the elevation is not set properly machine can not work normally Chapter 4 Operation Instruction I Basic Sterilization Illustration Operation...

Page 10: ...amber cover too fiercely so that it will not be extruded and broken 4 Loading articles to be sterilized 1 Put the articles to be sterilized into the baskets and close the cover 2 Make sure the water p...

Page 11: ...icator Locked is on The cover closed indicator may change repeatedly in course of tightening When it is nearly secured turn the handle slowly while observing whether the cover closed indicator is on o...

Page 12: ...1 When the set sterilizing time is up the system will send out a sound alarm and the steam will be exhausted to release automatically EXHT indicator starts to blink 2 Under solid mode with drying when...

Page 13: ...No sufficient water inside chamber makes heater dry scorch The elevation above 300m isn t set correctly on user administration menu Set correct elevation E04 Abnormal fluctuation of power results in a...

Page 14: ...ill not be used for a long period of time the water in the sterilizer chamber must be emptied to avoid blocking the pipeline 3 Wipe the surface of the sealing ring and its contact parts with soft and...

Page 15: ...tor 1 Under standby condition press and hold FUNC and DATA buttons then screen A will show 0000 Among these the first 0 flashing means it is changeable And press of UP or Down button may adjust the fi...

Page 16: ...de P011 Version of system unmodified P012 Solid mode exhaustion temperature setting 105 C 130 C Chapter 8 Diagram of Pipeline Interface 1 A series diagram of pipeline 1 Pressure gauge 2 Pressure switc...

Page 17: ...0A FD100A Electrical schematic diagram Drawell International Technology Limited Add Suite 1117 Lane561 XiuChuan Rd Greeland Max Mall PuDong Shanghai Tel 0086 21 54411195 Fax 0086 21 33823261 Web www d...