Precautions on Safe Operation

Precaution:

For your safety and to operate the instrument properly, please read this operation instruction

carefully before using and operate the following requirements in this manual.Where you violate
the instruction of manufacturer in this manual, it is possible to damage the protection of this
instrument.

This instrument shall not be used for purposes other than sterilization

drying,and it shall also

not be used for sterilizing the flammable, explosive and readily oxidizable products or strong acid,
alkali and saline solutions to avoid explosionand corruption of sterilizer chamber or pipelines.

The power supply of sterilizer should be properly connected according to nameplate on the

instrument. In case of excessive voltage fluctuation, the regulated power supply shall be used to
ensure optimum performance of the instrument. IF you are using other types of voltage, a
transformer must be used to prevent the instrument from being damaged.

It is necessary to ground the instrument properly, and never connect the ground wire to

plastic pipeline, gas pipe, telephone ground wire and lightning rod.

Never block the steam exhaust hole on the safety valve with articles, so that the safety valve

may work properly under abnormal conditions.

Always ensure that the reading of pressure gauge is "OMpa" before opening the chamber

cover. Never attempt to open the chamber cover and drainage valve, or draw out the steam
trapping bottle when the pressure inside the sterilizer chamber is over "OMpa" to avoid injury
from ejected pressurized steam.

Avoid dripping water to the control circuit when adding distilled water to the sterilizer

chamber to against electric shock or other faults.

When using waste processing bags or other type of bags, please put the bags into the

stainless steel basket before put into the sterilizer chamber to avoid affecting the precision of
temperature.

If the handle is turned by mistake during sterilization, it will not continue turning as it has

been locked by the interlocking device, and there will be a little steam exhausted to release. In this
case, press the"STOP" button to stop the instrument immediately. When opening the chamber
cover, turn the handle slightly toward the opposition direction that it is wrongly operated, and
then turn the handle counter-clockwise until the cover is released.

Observe the temperature variation inside the sterilizer chamber. The temperature is relatively

high when operation has been finished, keep the face and hands away from the sterilizer chamber
when opening the cover to avoid injury from ejected pressurized steam. Gloves must be used
when taking articles out of the sterilizer chamber.

Liquid does take some time to get cooling, verify whether the temperature is low enough or

not, when taking the liquid after sterilization out of the chamber to avoid scalding.

Summary of Contents for DW-FD Sereries

Page 1: ...Drawell Scientific USER S MANUAL DW FD Series FD Series Automatic Autoclave...

Page 2: ...oading articles to be sterilized 08 5 Closing the sterilizer chamber cover 09 6 Selecting sterilizing program 09 7 Starting sterilization 10 8 Opening the cover Program finished 10 9 Taking the steril...

Page 3: ...is OMpa before opening the chamber cover Never attempt to open the chamber cover and drainage valve or draw out the steam trapping bottle when the pressure inside the sterilizer chamber is over OMpa...

Page 4: ...peration to allow the instrument to cool down so that it will be able to produce adequate saturated steam In case of abnormal conditions such as noise odor and smog please turn off the power immediate...

Page 5: ...Drawell Scientific Chapter 1 External Features Controller I External Features...

Page 6: ...Drawell Scientific...

Page 7: ...sure gauge 0 0 4Mpa Working environment 5 40 C relative humidity 10 85 Rated working pressure 0 25Mpa Designed pressure 0 3Mpa Sterilization temperature 105 135 Sterilization timer 1 300min Drying tim...

Page 8: ...ake out the water plate remove the packaging place it in the bottom of the chamber in a horizontal position Connectioninstrument must be grounded reliably if the power socket does not 1 The instrument...

Page 9: ...to avoid affecting normal work of sterilization Caution If the elevation is not set properly machine can not work normally Chapter 4 Operation Instruction I Basic Sterilization Illustration Operation...

Page 10: ...amber cover too fiercely so that it will not be extruded and broken 4 Loading articles to be sterilized 1 Put the articles to be sterilized into the baskets and close the cover 2 Make sure the water p...

Page 11: ...icator Locked is on The cover closed indicator may change repeatedly in course of tightening When it is nearly secured turn the handle slowly while observing whether the cover closed indicator is on o...

Page 12: ...1 When the set sterilizing time is up the system will send out a sound alarm and the steam will be exhausted to release automatically EXHT indicator starts to blink 2 Under solid mode with drying when...

Page 13: ...No sufficient water inside chamber makes heater dry scorch The elevation above 300m isn t set correctly on user administration menu Set correct elevation E04 Abnormal fluctuation of power results in a...

Page 14: ...ill not be used for a long period of time the water in the sterilizer chamber must be emptied to avoid blocking the pipeline 3 Wipe the surface of the sealing ring and its contact parts with soft and...

Page 15: ...tor 1 Under standby condition press and hold FUNC and DATA buttons then screen A will show 0000 Among these the first 0 flashing means it is changeable And press of UP or Down button may adjust the fi...

Page 16: ...de P011 Version of system unmodified P012 Solid mode exhaustion temperature setting 105 C 130 C Chapter 8 Diagram of Pipeline Interface 1 A series diagram of pipeline 1 Pressure gauge 2 Pressure switc...

Page 17: ...0A FD100A Electrical schematic diagram Drawell International Technology Limited Add Suite 1117 Lane561 XiuChuan Rd Greeland Max Mall PuDong Shanghai Tel 0086 21 54411195 Fax 0086 21 33823261 Web www d...

Reviews: