Drawell Scientific
P003 Setting of pressure unit: 1: Kpa; 2: bar; 3: psi
;
4: No ( Not available or no operation
for pressure sensor)
P005: Printing language: 0: English 1: Chinese
P006: Safety valve test enabled or not (Yes: enabled No: disabled)
P007: Chamber temp sensor,temp compensation: the temp compensation scope is -5°
C - +5°C
P008:Setting of max working temp: the temp range is 135°C-145°C
P009: Setting of local altitude: (0-3000m)
P010: Boiling point temp corresponding to altitude( according to altitude)
P011: Version of system, unmodified
P012: Solid mode: exhaustion temperature setting (105°C-130°C)
Chapter 8 Diagram of Pipeline& Interface
1.A series diagram of pipeline
1. Pressure gauge 2. Pressure switch
3. Pressure sensor
4.Safety valve
5. One-way valve
6. Exhaust solenoid
7.Drainage valve
2.R series diagram of pipeline
1. Pressure gauge 2.Pressure switch 3.Pressure sensor 4.Safety valve
5.One-way valve 6.Exhaust solenoid
7. Drainage solenoid
8.Filter
9.Drainage valve 10. Drainage valve
3.FD65R/FD65A/FD80R/FD80A/FD100R/FD100A diagram of interface
4.FD36R/FD36A/FD50R/FD50A diagram of interface
Chapter 9 Wiring diagram
1.FD36R/FD50R/FD65R/FD80R/FD100R Electrical schematic diagram