Drawell DW-FD Sereries User Manual Download Page 13

Drawell Scientific

speed of liquid.

10.Turning off the power

1)Power must be switched off after the completion of daily operations or the instrument will not be used for a long period of
time.
2)When sterilization is completed of a day, it is recommended to drain out all water in sterilization chamber.

Chapter 5 Troubleshooting

1.The automatic control system of the sterilizer monitors the real time operation of the instrument. Whenever any failure
occurs, the system may send out alarm and display the error code, then, please press the STOP button to return to
standby status, and turn off the power, check the error code and handle on time.
2.To insure the safety, the chamber cover can be open under the circumstance that no pressure in the chamber and the
temperature is below 80

°C

at the same time

3.Under liquid mode

donot press STOP button before the temperature is lower than the boiling point,so as to avoid the

liquid loads overflowing.

Error Code

Possible Causes

Troubleshooting Method

E01

The cover of sterilizer chamber is not

secured

Turn the handle until the cover closed

indicator is on and the cover is

secured

E02

Heater is burnt

Contact the local dealer

E03

No sufficient water inside chamber

makes heater dry scorch.

The elevation above 300m isn’t set

correctly on user administration menu

Set correct elevation

E04

Abnormal fluctuation of power results in

abnormal change of temp

Whether the power supply is between

198V and 242V

E05

Solenoid

valve

jammed

by

contaminants leads to abnormal work of

exhaust valve

Contact the local dealer to change

the automatic steam exhaust value

E06

Temperature sensor comes off the

circuit board

Plug the temperature sensor in place

Temperature sensor is not functional

Replace the temperature sensor

E07

Plug or joint of temperature sensor is

short-circuited

Contact the local dealer to replace

the temperature sensor

E08

Temperature in the chamber when start

working is higher than 100°C

Wait until the temperature falls down

before restart

E10

Power failure in the course of

sterilization

Re-sterilize

E13

Failure of drying temp controller

Contact the local dealer

E14

Failure

of

heating

system

or

temperature control system

Contact the local dealer

E18

Over high pressure in sterilization

chamber

Stop working and contact the local

dealer

E20

Safety valve test circuit failed

Contact the local dealer

The above chart covers the simple problems, if you can not handle them by yourselves, please contact the local dealer
and provide the following information

1)Model and Number of Machine

Summary of Contents for DW-FD Sereries

Page 1: ...Drawell Scientific USER S MANUAL DW FD Series FD Series Automatic Autoclave...

Page 2: ...oading articles to be sterilized 08 5 Closing the sterilizer chamber cover 09 6 Selecting sterilizing program 09 7 Starting sterilization 10 8 Opening the cover Program finished 10 9 Taking the steril...

Page 3: ...is OMpa before opening the chamber cover Never attempt to open the chamber cover and drainage valve or draw out the steam trapping bottle when the pressure inside the sterilizer chamber is over OMpa...

Page 4: ...peration to allow the instrument to cool down so that it will be able to produce adequate saturated steam In case of abnormal conditions such as noise odor and smog please turn off the power immediate...

Page 5: ...Drawell Scientific Chapter 1 External Features Controller I External Features...

Page 6: ...Drawell Scientific...

Page 7: ...sure gauge 0 0 4Mpa Working environment 5 40 C relative humidity 10 85 Rated working pressure 0 25Mpa Designed pressure 0 3Mpa Sterilization temperature 105 135 Sterilization timer 1 300min Drying tim...

Page 8: ...ake out the water plate remove the packaging place it in the bottom of the chamber in a horizontal position Connectioninstrument must be grounded reliably if the power socket does not 1 The instrument...

Page 9: ...to avoid affecting normal work of sterilization Caution If the elevation is not set properly machine can not work normally Chapter 4 Operation Instruction I Basic Sterilization Illustration Operation...

Page 10: ...amber cover too fiercely so that it will not be extruded and broken 4 Loading articles to be sterilized 1 Put the articles to be sterilized into the baskets and close the cover 2 Make sure the water p...

Page 11: ...icator Locked is on The cover closed indicator may change repeatedly in course of tightening When it is nearly secured turn the handle slowly while observing whether the cover closed indicator is on o...

Page 12: ...1 When the set sterilizing time is up the system will send out a sound alarm and the steam will be exhausted to release automatically EXHT indicator starts to blink 2 Under solid mode with drying when...

Page 13: ...No sufficient water inside chamber makes heater dry scorch The elevation above 300m isn t set correctly on user administration menu Set correct elevation E04 Abnormal fluctuation of power results in a...

Page 14: ...ill not be used for a long period of time the water in the sterilizer chamber must be emptied to avoid blocking the pipeline 3 Wipe the surface of the sealing ring and its contact parts with soft and...

Page 15: ...tor 1 Under standby condition press and hold FUNC and DATA buttons then screen A will show 0000 Among these the first 0 flashing means it is changeable And press of UP or Down button may adjust the fi...

Page 16: ...de P011 Version of system unmodified P012 Solid mode exhaustion temperature setting 105 C 130 C Chapter 8 Diagram of Pipeline Interface 1 A series diagram of pipeline 1 Pressure gauge 2 Pressure switc...

Page 17: ...0A FD100A Electrical schematic diagram Drawell International Technology Limited Add Suite 1117 Lane561 XiuChuan Rd Greeland Max Mall PuDong Shanghai Tel 0086 21 54411195 Fax 0086 21 33823261 Web www d...

Reviews: