User Manual of CIC-D100 Ion Chromatograph
59
transportation. Finally, connect the exhaust pipe and sink it below the water surface.
:
The eluent generation tank is filled with KOH or MSA solution with strong
corrosivity. Please wear rubber gloves and goggles during operation to prevent burns.
6.11 Replacement of capture column
1. Turn off the power supply of the instrument;
2. Slowly lift the eluent generation tube to expose the capture column below;
3. Disconnect the cable of the capture column and remove the surrounding pipeline;
Figure 6-9 Capture column pipeline connection
4. Connect the capture column pipeline and cable;
5. Recover the eluent generation tank;
6. Power on the instrument to resume operation.
The eluent
generation tank OUT
to the capture
column IN and OUT
to the low-pressure
degassing box. The
connection sequence
can be reversed.
Conductor cell OUT
to capture column
REGEN IN. Capture
column REGEN
OUT to suppressor
REGEN IN. The
connection sequence
can be reversed.
Warning