19
Calling the Info-Off mode
With the instrument switched off, press the
[+]
key for approx. 2 seconds.
The name of the gas, measuring unit, and measuring range limit value are
displayed for all channels.
Pressing the
[+]
key again exits the Info Off Mode (or via timeout).
Calling the Quick Menu
In measuring mode, press the
[+]
key three times.
–
If functions in the quick menu are activated using the PC software "Dräger
CC Vision", you can select these functions using the
[+]
key. If no functions
have been activated in the quick menu, the instrument remains in measuring
mode. Possible functions:
–
Bump test (configuration for bump test, see technical manual)
–
Fresh air calibration
1)
–
Delete peak values
–
Display pump information, see technical manual
–
Activate or deactivate pump, see technical manual
Press the
[OK]
key to call the selected function.
Press the
[+]
key to cancel the active function and to switch to measuring
mode.
–
If no key is pressed for 60 seconds, the instrument returns automatically to
measuring mode.
1)
The fresh-gas adjustment/zero-point adjustment is not supported by the CO
2
channel of the
infrared sensor and the XXS O
3
. A zero-point calibration of these sensors can be made with the
Dräger CC-Vision PC software. A suitable zero gas free from carbon dioxide and ozone (e.g. N
2
)
must be used.
Replacing the batteries / rechargeable batteries
Switching off the instrument:
Press and hold the
[OK]
and
[+]
keys simultaneously.
Loosen the screw on the power pack and remove the power pack.
Battery holder (order no. 83 22 237):
WARNING
Explosion hazard! To reduce the risk of ignition of a flammable or explosive
atmosphere, strictly adhere to the following warning statements:
Do not throw used batteries into fire or try to open them by force.
Do not replace or charge batteries in potentially explosive areas.
Do not use new batteries with used batteries, and do not mix batteries from
different manufacturers or of different types.
Remove batteries before maintenance work.
Batteries / rechargeable batteries are part of the Ex approval.
Only the following types may be used:
–
Alkaline batteries – T3 – (non rechargeable!)
Panasonic LR6 Powerline
Varta Type 4106
1)
(power one) or
Varta Type 4006
1)
(industrial)
–
Alkaline batteries – T4 – (non rechargeable!)
Duracell Procell MN1500
1)
, Duracell Plus Power MN1500
1)
–
NiMH rechargeable batteries – T3 – (rechargeable)
GP 180AAHC
1)
(1800 mAh) max. 40 °C ambient temperature.
Only charge NiMH power packs T4 (type HBT 0000) or T4 HC (type HBT 0100)
with the appropriate Dräger charger. Charge NiMH single cells for ABT 0100
battery holder as directed by the manufacturer. Ambient temperature during
charging: 0 to +40 °C.
1)
Not part of the BVS10 ATEX E 080X and PFG 10 G 001X technical suitability test.
WARNING
Explosion hazard!
The Dräger X-am 5600 is only permitted to be operated with battery holder
ABT 0100 (X-am 5600), identified by the silver sticker.