16
Operation
Calling the Calibration Menu
— The calibration menu can only be ac-
cessed by entering a password.
Password on delivery:
» 001 «
— The default password on delivery can be
changed using the PC software "CC Visi-
on".
●
In measuring mode, press the
key for
at least 5 seconds.
— The function for entering the password is
selected.
— The special symbol »
« (for the "enter
password" function) is displayed.
— The display shows » 000 « with the first
digit flashing.
●
Use the
key to set the flashing digit.
●
Press the
key, the second digit starts
flashing.
●
Use the
key to set the flashing digit.
●
Press the
key, the third digit starts
flashing.
●
Use the
key to set the flashing digit.
●
Press the
key to confirm the password
once it has been set completely.
— The calibration menu functions can now
be selected by pressing the
key.
●
Press the
key to call the selected
function.
02323854_01.eps
OK
OK
OK
OK