26
Faults, Cause and Remedy
202
The span calibration of the
DrägerSensor CatEx 125 PR
is not valid
Carry out the span calibration,
203
The measurement value of the
DrägerSensor CatEx 125 PR
is in the negative range
Carry out the fresh air calibra-
204
The DrägerSensor CatEx
125 PR is not inserted
Check the DrägerSensor CatEx
125 PR, page 46.
205
Error during the function test
with gas (bump test) of the
DrägerSensor CatEx 125 PR
Repeat the function test, cali-
brate or replace the
DrägerSensor CatEx 125 PR, if
necessary, page 46.
206
Function not possible:
O
2
deficiency
Use DrägerSensor CatEx
125 PR in atmospheres with
more than 10.0 % O
2
by vol..
207
Failed filter test.
Repeat filter test with X-dock.
208
Failed rise time test.
Repeat rise time test with X-dock.
301
The zero point calibration of
DrägerSensor XXS O
2
is not
valid
Carry out the fresh air calibra-
302
The span calibration of the
DrägerSensor XXS O
2
is not
valid
Carry out the span calibration.
page 40 or carry out the fresh air
calibration, page 36.
303
O
2
concentration drifted into
the negative range
Fresh air calibration required,
304
The DrägerSensor XXS O
2
is
not inserted
Check the DrägerSensor XXS
O
2
305
Error during the function test
with gas (bump test) of the
DrägerSensor XXS O
2
Repeat the function test, cali-
brate or replace the DrägerSens-
or XXS O
2
if necessary, page 46.
307
Failed filter test.
Repeat filter test with X-dock.
308
Failed rise time test.
Repeat rise time test with X-dock.
401
The zero point calibration of
DrägerSensor XXS H
2
S is not
valid
Carry out the fresh air calibra-
402
The span calibration of the
DrägerSensor XXS H
2
S is not
valid
Carry out the span calibration,
Special symbol
»
«
and displayed numerical
code:
Cause
Remedy