
10
Configuration
Configuration
Standard Gas Configuration
Standard Device Configuration on Delivery
Changing the standard configuration: See “Configuring the Device” on page 21.
___________________
1)
In the case of O
2
A1 is the lower alarm setpoint: an alarm is issued if the value is too low.
2)
Different settings can be selected to meet customer requirements on delivery. The current setting can
be checked and changed with the software "CC Vision".
A version of Dräger CC-Vision which can be used for Dräger X-am 1700/2000 is supplied with the
device on CD.
DrägerSensor
CatEx
125 PR
[%LEL]
XXS O
2
[Vol.-%]
XXS CO
[ppm]
XXS H
2
S LC
[ppm]
Measuring range
2)
0 to 100
0 to 25
0 to 2000
0 to 100
Alarm A1
2)
- setpoint
- can be acknowledged
- self-latching
20
yes
no
19
1)
no
yes
30
yes
no
10
yes
no
Alarm A2
2)
- setpoint
- can be acknowledged
- self-latching
40
no
yes
23
no
yes
60
no
yes
20
no
yes
X-am 1700 / X-am 2000
Function test with gas (bump test) in
Quick Menu
2)
Quick bump test
Fresh air calibration in Quick Menu
2)
off
Life sign
2)
on
Switch off
2)
allowed
LEL factor
2)
(CH
4
)
4.4 (4.4 vol. % corresponds to 100 %LEL)
Averaging time
2)
15 minutes for STEL
8 hours for TWA
WARNING
After a basic initialization has been carried out with the PC software Dräger CC Vision, indi-
vidual alarm settings may have been changed.