Dräger X-act 5000 Basic
Maintenance
4.2
Warnings
Warnings will be displayed together with the corresponding icon, and
if necessary, together with a written hint.
Possible warnings include:
Mandatory leak test
Activation strokes required
Desorption strokes required
No measurement in explosion-hazard areas
Battery empty
Invalid password
Replace SO
3
filter
"PUMPE <5°C/41°F START?"
5
Maintenance
5.1
Maintenance table
104
Boot loader
communication fault.
Contact DrägerService.
105
Motor does not start.
Contact DrägerService.
106
Motor does not stop.
Contact DrägerService.
107
Motor too slow.
Contact DrägerService.
108
Motor too fast.
Contact DrägerService.
109
APS sensor out of
range.
Contact DrägerService.
110
DPS sensor out of
range.
Contact DrägerService.
111
DPS sensor fault
Contact DrägerService.
112
Software integrity fault Contact DrägerService.
113
Data integrity fault
Contact DrägerService.
114
Volume out of range.
Calibrate flow or
contact DrägerService.
Text/Code
Cause
Remedy
WARNING
To prevent ignition of flammable or combustible atmospheres,
disconnect the power supply before carrying out maintenance
work on the Dräger X-act 5000 Basic.
Work to be performed
Eve
ry
year
E
ver
y t
w
o
y
ears
Calibrate flow.
X
Have the Dräger X-act 5000 Basic inspected by Dräger.
X
Change SO
3
filter.
X
1)
1)
Or if indicated on the display
!