Supplement Scio Four modules
Connector
9-pole Sub-D, galvanically isolated 1.5 kV against internal electronics, 0.5 kV
against housing.
Baud rate
1200, 2400, 4800, 9600, 19200, 38400 baud
Data bits
8
Parity
Even
Stop bits
1
Pin assignment
Pin 1, 4, 6, 7, 8, 9
Open
Pin 2
TXD
Pin 3
RXD
Pin 5
GND
Housing
SHLD
Gas measurement
Method
Sidestream gas measurement
All values are measured under ATPS conditions, sampling rate calibrated under
NTPD conditions.
The endtidal gas concentration is the value measured at the time of maximum CO
2
concentration in the expiratory phase.
Sampling rate
200 mL/min ±20 mL/min
Maximum time to empty water trap
41 hrs (sample gas under BTPS conditions, ambient air 23 °C)
Gas leakage
<10 mL/min
Data sample rate
<50 ms
Delay time
<2.4 s
Total system response time
<3 s
Sum of delay time and gas-specific rise time (T10...90)
Cross sensitivity
<0.2 Vol% for volatile agents, CO
2
, and N
2
O
<1 Vol% for O
2
against ethanol (<3000 ppm), isopropanol (<10000 ppm), acetone (<1000 ppm),
methane (<1000 ppm), saturated humidity at 37 °C
Drift
Compensated by automated cyclical zeroing
O
21)
2)
Range
0 to 100 Vol%
Accuracy
±(2.5 Vol% + 2.5 % rel.)
Rise time (T10...90)
<500 ms
CO
21) 2)
Range
0 to 10 Vol%
Accuracy
±(0.43 Vol% + 8 % rel.)
Rise time (T10...90)
<300 ms
N
2
O
1) 2)
Range
0 to 100 Vol%
Accuracy
±(2 Vol% + 8 % rel.)
Rise time (T10...90)
<300 ms
Anesthetic agents
1) 2)
Range
Halothane
0 to 8.5 Vol%
Isoflurane
0 to 8.5 Vol%
Enflurane
0 to 10 Vol%
Sevoflurane
0 to 10 Vol%
Desflurane
0 to 20 Vol%
Accuracy
±(0.20 Vol% + 15 % rel.)
Summary of Contents for Scio Four
Page 19: ...Supplement Scio Four modules 19 This page intentionally left blank...
Page 23: ...Scio Four 23 MC IEC 60601 1 2 MC MC MC MEDIBUS 0 3 1 0 MEDIBUS 38 Dr ger...
Page 27: ...Scio Four 27 1 RS232C X2 A 2 Infinity 1 HME Luer Lock 2 1 2 002 A...
Page 28: ...28 Scio Four 1 2 1 2 3 1 1 2 HME...
Page 30: ...30 Scio Four 1 2 3 4 1 2 3 4 Dr ger Dr gerService 1 1 12 10...
Page 39: ...Scio Four 39...
Page 129: ...Dodatak Scio Four moduli 129 Ova je stranica namjerno ostavljena prazna...
Page 151: ...Scio Four 151 MEDIBUS Scio Four CO2 N2O O2 0 3 m 1 0 MEDIBUS 166 Dr ger PVC...
Page 155: ...Scio Four 155 1 Y Luer Lock 2 1 2 AGSS 1 2 O 1 2...
Page 156: ...156 Scio Four 1 2 3 1 Dismozon plus Bode Chemie...
Page 163: ...Scio Four 163 CISPR 11 1 A IEC 61000 3 2 IEC 61000 3 3...
Page 167: ...Scio Four 167...
Page 171: ...Scio Four 171 MEDIBUS Scio Four CO2 N2O O2 0 3 m 1 0 ft MEDIBUS 186 Dr ger PVC...
Page 172: ...172 Scio Four 25 MRI NMR NMI 25 cm 10 in Dr ger 001 B 002 B MEDIBUS H RS232C X2 B B H...
Page 175: ...Scio Four 175 1 Luer Lock 2 1 2 1 2...
Page 176: ...176 Scio Four 1 2 1 2 3 1...
Page 187: ...Scio Four 187...
Page 205: ...Ek Scio Four mod lleri 205 Bu sayfa kasten bo b rak lm t r...
Page 212: ...212 Scio Four 1 Y HME Luer 2 1 2 1 On Off ON 2 O 1 2...
Page 229: ...Scio Four 229 1 ON OFF ON 2 1 2 3 OFF 1 ON OFF OFF 1 2 HME...
Page 239: ...Scio Four 239...