Technical data
Dräger Polytron 8100
43
10.2 Marking
The marking is reproduced on a separate piece of paper
shipped with the instrument.
Serial Number key: The third letter of the serial number
specifies the manufacturing year: A = 2009, B = 2010,
C = 2011, D = 2012, E = 2013, F = 2014, H = 2015, J = 2016,
K = 2017, etc.
Example: Serial Number ARCB-0103: the third letter is C,
which means that the unit was manufactured in 2011.
10.3 Signal transmission to central controller
10.3.1 Analog signal
10.4 Power supply and relays
NOTICE
The table shows the response of the sensor to other
gases than the target gas (cross sensitivities). The
listed values are typical and are valid for new sensors.
The table does not claim to be complete. Gas mixtures
can be displayed as the sum of all components.
Gases with a negative cross sensitivity can cancel out
a positive sensor signal. The influence of the O
2
displacement is not taken into account (partial
pressure measurement).
Example:
2 Vol% ethylene in the air - cross sensitivity (table
value) = -1 Vol% O
2
.
O
2
displacement caused by 2 Vol% ethylene (2% of 20
Vol% O
2
) = -0.4 Vol% O
2
Instrument display (20.9 - 1 - 0.4) = 19.5 Vol% O
2
.
CAUTION
The exposure to higher concentrations of unsaturated
hydrocarbons, alcohols or hydrogen over an extended
period of time (dose approx. 100,000 ppm x hours)
could cause the sensor to fail.
i
i
!
NOTICE
Frequency of measurement calculation: 1 x per
second (update of display, analog interface and
relays).
Description
Analog output
Normal operation
4 to 20 mA
Drift below zero
3.8 to 4 mA
Measuring range exceeded
20 to 20.5 mA
Instrument fault
≤
1.2 mA
Fault on analog interface
> 21 mA
Maintenance signal
3.4 mA steady signal or 1 Hz
modulation between 3 and
5 mA (selectable)
Operating voltage
10 to 30 V
at the instrument
Cable specification
3-core shielded cable
Inrush current
2.3 A for 2 ms at 24 V
,
10
Ω
resistor
Operating current
(max.)
80 mA at 24 V
, without relay,
non-remote sensor
100 mA at 24 V
, with relay,
remote sensor
Relay rating (option)
SPDT, 0.1 A - 5 A at 230 V~,
0.1 A (minimum) - 5 A at 30 V
,
resistive load.
For safety-related applications (SIL 2),
the maximum contact rating is reduced
(see Dräger Polytron 8x00 Safety
Manual).
i
i
Summary of Contents for Polytron 8100
Page 2: ......
Page 3: ...1 5 4 3 2 5 1 4 3 10 7 8 9 2 6 3 4 5 2 1 3 9 5 2 1 14 13 3 4 5 6 7 8 14 2 11 12 15 A B C D E...
Page 223: ...Dr ger Polytron 8100 223 4 5 2 4 5 3 OK OK 4 5 4 OK 5 OK 3 3 2 2...
Page 250: ...250 Polytron 8100 Additional Approvals A 3 Control Drawings...
Page 252: ...252 Polytron 8100 Additional Approvals A 4 Declaration of conformity...
Page 253: ......