Operation
58
Instructions for use
Polaris 600
Replacing the battery in the Polaris 600 remote control for MedView
camera
The remote control requires 9-volt batteries of the
type 6LR61, 6F22, or 1604D.
The operating time of the remote control depends
on the capacity of the battery used and is
approximately 1 year for the factory-supplied
battery.
1
Using light pressure, raise the housing flap (A)
at the hinge at the bottom while at the same
time folding it out at the top.
2
Unlock the cover (B) with a ballpoint pen or
screwdriver and fold it out.
3
Withdraw the battery (C) together with its clip-
on connector (D) from the compartment.
WARNING
Risk of damage to the device
Batteries must be stored in accordance with
the manufacturer's specifications. Batteries
must not be disassembled, opened or
crushed. Batteries must not be short-
circuited. Batteries must not be exposed to
either heat or fire. Batteries must not be
stored in direct sunlight.
CAUTION
Risk of damage to the device
Only 6LR61, 6F22, or 1604D battery types may
be used, as otherwise the operation of the
remote control may be compromised.
CAUTION
Risk of damage to the device
Non-rechargeable batteries must not be charged.
CAUTION
Risk of damage to the device
The battery must be removed if the remote
control will not be used for a certain period of
time. If there is fluid escaping from the battery,
the fluid must not come into contact with the skin
or get into the eyes. If there has been contact,
rinse the affected area with plenty of water and
seek medical help immediately.