Dräger PAS AirPack 1 Trolley Module
Airline Compressed Breathing Air Supply System
Instructions for Use
3352691 : 'D' Edition : October 2003 : Subject to Modification
1
1642
Approvals
The Dräger PAS AirPack 1 Trolley Module conforms to the requirements of EN139 and meets the
requirements of EC Council Directive PPE (89/686/EEC) and PED (97/23/EC).
PPE EC Type examiner;
SGS United Kingdom Limited, Unit 202b, Worle Parkway, Weston-super-
Mare, BS22 0WA, UK. Notified Body No. 0120. CE Mark approved and issued with EC Type Examination
Certificate.
PED EC Type examiner;
Lloyds Register, Hiramford, Middlemarch, Office Village, Siskin Drive, Coventry,
CV3 4FJ, England. Notified Body No. 0038. The equipment is manufactured in accordance with the
requirements of Conformity Assessment Trolley Modules B + D of the Pressure Equipment Regulations.
A copy of the Declaration of Conformity is available from Dräger on request.
For Your Safety
The Dräger PAS AirPack 1 Trolley Module incorporates a high performance pressure reducer. The reducer
is factory set and must not be tampered with, or removed from the manifold, as this invalidates the
conditions of the Dräger warranty. Correct operational condition is valid only if Dräger service and re-
seals the pressure reducer.
u
Use of this equipment requires wearer training and observance of these Instructions for Use and the
Instructions for Use supplied with the selected associated breathing equipment.
u
Use the equipment for the purpose specified in this instruction, or as confirmed in writing by Dräger.
u
Use and maintenance of this equipment requires knowledge and compliance with National Regulations,
Laws and Standards governing the use of respiratory equipment in the country of use.
u
Only trained competent personnel should inspect and service the equipment at regular intervals and
a record kept of such inspections and service.
u
Only trained and competent personnel should carry out the charging of the compressed air cylinders.
u
Dräger recommends that a Service Contract be obtained from your Dräger Branch or Agent.
u
Contact Dräger for details of Service Contracts and Service Training Courses.
u
Use only original Dräger Spare Parts for service and maintenance.
u
Use only original Dräger Test Equipment for service and maintenance.
u
Notify Dräger if there is a component fault or failure.
Liability Statement
Terms and Conditions of warranty for the Dräger PAS AirPack 1 Trolley Module can be obtained from Dräger
on request. Responsibility for reliable function of the equipment transfers to the owner or operator when
serviced or repaired by untrained personnel, (not employed or authorised by Dräger), or when used in a manner
not conforming to its intended use.
Description and Intended Use
The Dräger PAS AirPack 1 Trolley Module is a
compact mobile two-wheel trolley unit with
flexible rubber airline hose and hose winding drum,
plus storage capacity for two valved high-pressure
cylinder assemblies. The module provides a low-
pressure independent breathing air supply to the
wearer of approved Dräger breathing apparatus
fitted with airline manifold connections.
The equipment is compatible with a range of
associated breathing apparatus and
accessories and must be used only with air that
complies with the air quality requirements of
respirable breathable air at the pressure and
quantities (volumes) required. When the airline
fitting of the breathing apparatus is connected
to the module, the approved configuration
provides the wearer with respiratory protection
when working in a contaminated or oxygen
deficient gaseous atmosphere.
The manifold assembly of the module allows for
the repeated removal and fitting of cylinders,
transferring of air supply from one cylinder to
another. This feature provides a means of
providing an uninterrupted air supply to the
wearer.
Automatic activation of a preset mechanical high
pressure whistle warning unit (HPWWU) alerts
the Controller of low cylinder pressure. Activation
of a low pressure whistle warning unit (LPWWU)
alerts the Controller of a drop in pressure of the
low-pressure air supply.
If required, the cylinders may be bypassed by
connecting the inlet coupling of the trolley
module to an independent low-pressure air
supply, e.g. from a works airline or LP
compressor (refer to Main Airline Pressures and
Flow and Air Quality requirement . A drop in
low air supply pressure is highlighted to the
Controller by the activation of an audible low
pressure whistle warning unit (LPWWU).
The possible available extended working duration,
depends on the configuration selected (attached to
works airline or LP compressor - with filter unit to
maintain air quality), the capacity (volume) of
cylinder(s) selected, the number of fully charged
cylinders available and the breathing rate of the
wearer(s).
Important Note: No more than two users of
approved airline breathing equipment
should be connected to the AirPack 1
Trolley Module at any time, and in any
configuration. Refer to the Equipment
Configuration details chart.
Details of equipment variants, associated
breathing equipment and accessories are
available from Dräger on request.
Technical Data
High Pressure Connections
Standard G5/8 as per
EN 144-2
.
200bar
or
300bar.
Other high-pressure (HP) connections are
available to National Standards.
Operating Pressure
Low Pressure -
8bar
Nominal. (
6bar to 10bar
)
Whistle Warning Units (WWU)
HPWWU - Preset between
60bar
to
50bar
LPWWU - Preset between
3bar
to
5bar
Compressed Air Cylinders
Contact Dräger for details of approved valved
compressed air cylinder assemblies available for
use with this equipment.
Important Note: Only matched pairs of cylinders
should be fitted to the Dräger AirPack 1
Trolley
Module, e.g. 2 x 9 litres 200bar. Do Not mix
200bar and 300bar cylinders.
Main Airline Pressures and Flow
Safety Note: In order for the Trolley Module to
function efficiently, it is the responsibility of
the user to ensure that the following air
pressure and flows are available from the
factory mains airline air source. Air quality
for compressed air breathing systems must
conform to the requirements of EN12021.
The selected LP independent air supply must meet
the following parameters.
u
One User
7bar to 10bar - airflow rate of at
least 550 Litres/minute.
u
Two Users
7bar to 10bar - airflow rate of
at least 550 Litres/minute.
LP Leak Test
- Indicated pressure should not
decrease more than
1bar
in one minute.
Preparation for Use
Only trained and competent personnel must
perform the following checks and preparation
procedures before release of the equipment for
operational use to a 'Controller' and potential
breathing apparatus 'Wearer'.
Visual Inspection
Check the cleanliness and integrity of,
u
Trolley frame and wheels.
u
Cylinder support straps (2).
u
Winding Drum and low pressure hose.
u
Pneumatics, Gauges, WWUs, Quick Release
Coupling(s) (QRC) etc.
u
HP hoses and cylinder connector handwheel
assemblies (2).
Fitting the Cylinders
u
Place the AirPack 1 Trolley Module in the upright
position as shown in Fig. 1.
u
Check that the valve outlet port of each fully
charged valved cylinder is clean and
undamaged. Place each cylinder through the
loop of each of the cylinder straps locating the
rounded end of the cylinders into the cylinder
cradle supports at the base of the trolley frame.
u
Check that both connector handwheels are not
damaged and that the connector O rings are
in position and not damaged.
u
Rotate the cylinders to align the outlet port of
each valve with the associated connector
handwheel assembly. Ensure a smooth curve
of the high-pressure hose. Screw each
handwheel into the appropriate cylinder valve
hand tight.
u
Take up slack of each of the cylinder support
straps then activate the Camlock mechanism
by pulling the free end of the support strap
around the cylinder - securing each cylinder
to the frame. Fig. 1.
Leak Test and Test of WWUs
u
'Close'
both the high-pressure (HP) bleed
valves (6) of the cylinder connector
handwheel assemblies - handtight.
u
'Open'
slowly, but fully,
ONE
of the cylinder
valves to pressurise the system. Check the
indicated pressure readings on the HP gauge
(2) and the LP gauge (3).
High Pressure
- Indicated pressure should be
at least
80%
of maximum working pressure
indicated on the selected cylinders.
Low Pressure
- Indicated pressure should be
between
6bar
to
10bar
.
u
'Open'
the HP bleed valve (6) of the connector
handwheel assembly of the opposite cylinder
(i.e. the Closed valve) there should be no
audible leak. If satisfactory
'Close'
the HP
bleed valve.
u
'Close'
the cylinder valve.
Wait for 10 to 15
seconds to allow the system to stabilise then
c
heck the pressure reading on the HP gauge
(2) and the LP gauge (3).
HP Leak Test
- Indicated pressure should not
decrease more than
20bar
in one minute.
u
Following satisfactory leak test, check the
activation of the HPWWU (1) and the LPWWU
(4) as follows:
Slowly vent pressure from the HP side of the
system by gripping and carefully pulling the
handle of the relief valve (7). As the pressure
decreases observe the HP gauge (2) .
HPWWU
- the whistle should begin to sound
between
60bar
to
50bar
. As soon as the whistle
sounds, release grip on the handle of the relief
valve (7) and allow the whistle to vent. When
the HPWWU stops sounding proceed
immediately to test the LPWWU.
u
Following satisfactory HPWWU test, check the
activation of the LPWWU as follows:
Again grip and carefully pull the handle of the
relief valve (7). As the pressure decreases
observe the LP gauge (3) .
LPWWU
- the whistle should begin to sound
between
5bar
to
3bar
.
u
Check the HP bleed valves (6) at each cylinder
valve are
'Closed'
then slowly, but fully,
'Open'
the
second
cylinder valve to pressurise the
system. Check the indicated pressure readings
on the HP gauge (2) and the LP gauge (3).
High Pressure
- Indicated pressure should be
at least
80%
of maximum working pressure
indicated on the selected cylinders.
Low Pressure
- Indicated pressure should be
between
6bar
to
10bar
.
u
Open the high-pressure bleed valve (6) of the
connector handwheel assembly of the opposite
(first) cylinder (i.e. the closed valve.) there
should be no audible leak. If satisfactory
'Close'
the bleed valve.
u
'Close'
the cylinder valve and then check the
pressure reading on the HP gauge (2).
High Pressure Leak Test
- Indicated pressure
should not decrease more than
20bar
in one
minute.
u
Following satisfactory completed tests, grip and
pull the handle of the relief valve (7) to vent
pressure from system.
u
Having passed the tests, the equipment is now
available for use.
Pre-Use Procedures
Caution: Before connecting the selected airline
breathing equipment to the tested and released
Dräger
PAS AirPack 1
Trolley Module, check
that the breathing equipment has been tested
by a trained and competent person before
being released for use. When in use the
AirPack 1
Trolley Module must be positioned
in a safe and uncontaminated area. Dräger
recommend that a Controller must be present
at all times to monitor and maintain the air
supply to the wearer of the selected breathing
equipment. Time required for the wearer of
the airline breathing equipment to safely
evacuate from the hazardous area to a safe area
must be within the air capacity of the remaining
cylinder(s) of the
PAS AirPack 1
Trolley Module.
Dräger recommend that the user(s) of this final
configuration operate and maintain a Permit
for Use procedure and also have an
established company evacuation procedure(s).
The Dräger PAS AirPack 1 Trolley Module may be
used either as a self-contained unit, i.e. using the
attached compressed air cylinders, or by connecting
the module to an independent low-pressure air supply,
e.g. from a works airline or LP compressor the
cylinders can be bypassed.
Self-contained Unit
The Controller
u
With the trolley module suitably positioned in a
safe and uncontaminated area, pull and unwind
a required length of hose from the winding drum.
Attach any required accessories, e.g. extension
hoses, 'Y' pieces, etc.
u
Check that both HP bleed valves (6) at each of
the connector handwheel assemblies are
'Closed'
(clockwise).
When used as Self-contained Unit
u
'Open'
slowly, but fully,
ONE
of the cylinder
valves to pressurise the system. Check the
indicated pressure readings on the HP gauge
(2) and the LP gauge (3).
High Pressure
- Indicated pressure should be
at least
80%
of maximum working pressure
indicated on the selected cylinders.
Low Pressure
- Indicated pressure should be
between
6bar
to
10bar
.
When used from a works airline or
LP compressor
Safety Note: Refer to Airline Pressure and
Flow. Air quality must conform to the
requirements of EN12021. If required use the
Dräger Filter Unit.
u
Connect an approved extension hose from the
selected LP independent air supply to the male
connector (5) of the trolley manifold.
'Open'
the valve of the independent air supply to
pressurise the system. Check indicated
pressure reading on the LP gauge (3).
Low Pressure
- Indicated pressure should be
between
6bar
to
10bar
.
The Controller
Important Note: Should the 'Controller' be
required to wear respiratory protection
equipment then all the following 'Wearer'
instructions should be observed. Connect the
male coupling of the selected breathing
equipment to the female QRC (8) of the pressure
manifold. This will reduce the duration available
from the selected system. It must be observed
however that no more than two users of
approved airline breathing equipment should
be connected to the PAS AirPack 1 Trolley
Module at any time, and in any configuration.
The Wearer
u
Refer to the Instruction for Use for the selected
breathing equipment, put on the equipment and
prepare the equipment for attachment and use
with the PAS AirPack 1 Trolley Module.
u
Connect the male coupling of the breathing
equipment to the female QRC of the supply
hose from the trolley module, i.e. the winding
drum.
Safety Warning: Facial hair, beard stubble,
side-whiskers, and the wearing of spectacles
will adversely affect and interfere with
facepiece seal. Correct fit of facepiece only
ensured if facepiece seal makes close contact
with skin.
u
Put on the facepiece as defined in the Instruction
for Use for the selected breathing equipment,
breathe normally and perform Function Check.
Function Check
The Controller
u
'Close'
the cylinder valve or the valve of the
associated air supply.
The Wearer
u
Breathe normally to vent air from the system.
During venting the whistle(s) will alarm at the
preset pressure. When fully vented, hold
breath. The facepiece should hold onto the face
indicating a positive seal.
u
Immediately indicate to the Controller of a
positive face seal and request to immediately
Open
the valve of the associated air supply.
The Controller
u
When Wearer indicates to the Controller of a
positive face seal the Controller should
immediately
Open
the associated valve to
pressurise the system.
The Wearer In Use
u
Inhale and hold breath. The unit must balance,
i.e. no audible leak. Continue to breathe normally.
Expired air should flow easily out of the exhalation
valve of the facepiece. Press the centre of the
rubber cover of the LDV to check for
supplementary supply airflow then release.
u
Continue to breathe normally and as instructed
by the controller proceed to the working area
taking care to support and route the airline hose
as necessary.
In Use - When used as a Self-
contained Unit
The Controller
u
Regularly take a HP gauge reading. At between
60bar
to
50bar
the HPWWU unit sounds
indicating that the cylinder in use is almost
empty. To therefore maintain the air supply to
the wearer of the breathing equipment
immediately proceed as follows:
1
. Slowly, but fully,
'Open'
the valve of the
second
fully charged cylinder. The HP gauge
(3) will indicate an increase in system pressure.
The wearer is now connected to the fully
charged cylinder.
2
.
'Close'
the valve of the empty cylinder then
'Open'
the bleed valve (6) of the associated
valve to vent the remaining pressure from the
HP supply hose.
3. Unscrew the connector handwheel from the
valve of the depleted cylinder. Release the
Camlock mechanism of the cylinder support
strap and remove the cylinder.
4. Select an available replacement fully charged
cylinder and assemble to the trolley. See -
Preparation for Use Fitting Cylinders.
5.
'Close'
the bleed valve (6) of the connector
handwheel assembly of the replacement
cylinder.
Note:
To maintain a supply of respirable breathing
air to the wearer of the breathing equipment, this
procedure can be repeated with further available
replacement fully charged cylinders.
Safety Warning: Should the LPWWU of the PAS
AirPack 1 Trolley Module sound, then the
controller must immediately implement the
agreed and established company evacuation
procedure(s).
The Wearer Self Contained Operation
u
Breathe normally.
u
When the task is completed, or when advised
by the Controller, proceed to the safe area
(Controller) by the safest route.
Important Note: - Refer to the evacuation
procedures as defined in the Instructions for
Use for the selected breathing equipment.
In Use - When used from a
Works Airline or LP Compressor
Safety Note: Refer to Airline Pressure and Flow.
Air quality must conform to the requirements
of EN12021. If required use the Dräger Filter
Unit.
The Controller
u
Take regular LP gauge (3) reading. The indicated
pressure should be between
6bar
to
10bar
.
Safety Warning: Should the LPWWU sound,
indicating a reduction in pressure from the
Works Airline or LP Compressor, the Controller
must immediately fully 'Open' ONE of the
cylinder valves of the
PAS AirPack 1
Trolley
Module to pressurise the system. The extension
hose from the selected LP independent air
supply to the male connector (5) should be
disconnected.
u
Continue to follow the instructions outlined in
the section i.e.
In Use- When used as Self-
contained Unit.
The Wearer - Works Airline etc.
u
Breathe normally.
u
When the task is completed, or when advised
by the Controller, proceed to the safe area
(Controller) by the safest route.
Important Note: - If necessary refer to the
evacuation procedures as defined in the
Instructions for Use for the selected breathing
equipment.
After Use
The Wearer
u
Refer to the After Use instructions in the
appropriate Instruction for Use for the selected
breathing equipment. Remove the equipment
and detach the male coupling of the breathing
equipment from the female QRC of the supply
hose of the carrying frame module.
Safety Warning: Do Not attempt to remove the
breathing equipment until in safe area, clear of
the hazard.
The Controller
u
Following the removal of the breathing
equipment by the wearer either:
1.
disconnect the extension hose from the
selected LP independent air supply from the
male connector (5).
or
2.
'Close'
the relevant cylinder valve of the
trolley module.
u
Grip and pull the handle of the relief valve (7) to
vent pressure completely from system.
'Open'
the (HP) bleed valves (6) of the cylinder
connector handwheel assemblies
u
Carefully and neatly rewind the airline hose
to the winding drum. Dräger recommend that
any dirt or contaminants on the hose should
be removed as it is being rewound onto the
winding drum.
Routine Maintenance
To be performed after use of the equipment.
See also Maintenance and Test Intervals Chart.
Refer to the Instructions for Use supplied with the
selected breathing equipment, facepiece, lung
demand valve etc.
Visual Inspection
Check the cleanliness and integrity of,
u
Trolley frame and wheels.
u
Cylinder support straps (2).
u
Winding Drum and LP hose.
u
Pneumatics, Gauges, WWUs, Quick Release
Coupling(s) (QCR) etc.
u
HP hoses and cylinder connector handwheel
assemblies (2).
u
Associated extension hoses.
Cleaning, Disinfecting, Drying
Dräger recommend that contaminated and dirty
components or assemblies, hoses etc. are
carefully cleaned and disinfected then thoroughly
dried after use, or as considered necessary by the
user.
To ensure correct operational condition of the
equipment use the cleaning and disinfecting
solutions recommended by Dräger. The use of
any other product will invalidate the Dräger
warranty and guarantee.
Safety Note: Refer to manufacturers usage
instructions when using cleaning and
disinfecting agents. It is important that
attention be paid to concentration and reaction
times. Do Not use organic solvents, such as
Acetone,
Alcohol,
White
Spirit,
Trichloroethylene or similar.
Dräger recommends the following:
1. Cleaning
u
Dräger Safety Wash.
u
Sekusept
Do Not
exceed a temperature of
30 degrees
Celsius.
Note:
Before disinfecting, rinse off cleaning
solution in clean water.
2. Disinfecting
u
Incidur
Do Not
exceed a temperature of
30 degrees
Celsius.
Note:
Before drying, rinse off
disinfecting
solution
in clean water.
u
Dräger Wipex Cloths.
Details of cleaning and disinfecting agents are
available from Dräger on request.
3. Rinsing and Drying
Remove cleaning and disinfecting solutions by
rinsing in clean water, followed by drying.
Do Not
exceed a temperature of
60 degrees
Celsius when drying components.
Charging Cylinders
Refer to 'For Your Safety'
Safety Warning: Air quality for compressed air
breathing systems must conform to the
requirements of EN12021.
Only charge compressed air cylinders which:
u
Conform to National Standards.
u
Feature the original manufacturers test date
and test mark.
u
Have not exceeded the test date indicated
on the cylinder by the last testing station.
u
Do Not used damaged cylinders.
Recharge to the rated pressure indicated on the
label or stamped on the neck or shoulder of
cylinder. Dräger recommend a charge rate of
27bar/minute. Rapid charging will induce an
increase in temperature resulting in an incomplete
charge - check pressure at ambient and if required
Top Up charge.
To prevent overcharging of the selected cylinder
Dräger recommend that a pressure-limiting device
is fitted to the charging compressor.
Storage - Ready for Use
u
Store the Trolley Module in an upright position
as shown in Fig. 1. A protective dust cover is
available. Contact Dräger for details.
u
Store the equipment Ready for Use in a cool
dry environment, free from dust and dirt.
Protect rubber parts -
Do Not
expose to direct
sunlight.
7
5
6
3
2
4
1
8