background image

Dräger PAS AirPack 1 Trolley Module

Airline Compressed Breathing Air Supply System

Instructions for Use

3352691 : 'D'  Edition : October 2003 : Subject to Modification

1

1642

Approvals

The Dräger PAS AirPack 1 Trolley Module conforms to the requirements of EN139 and meets the

requirements of EC Council Directive PPE (89/686/EEC) and PED (97/23/EC).

PPE – EC Type examiner;

 SGS United Kingdom Limited, Unit 202b, Worle Parkway, Weston-super-

Mare, BS22 0WA, UK.  Notified Body No. 0120.  CE Mark approved and issued with EC Type Examination

Certificate.

PED – EC Type examiner;

 Lloyds Register, Hiramford, Middlemarch, Office Village, Siskin Drive, Coventry,

CV3 4FJ, England.  Notified Body No. 0038.  The equipment is manufactured in accordance with the

requirements of Conformity Assessment Trolley Modules B + D of the Pressure Equipment Regulations.

A copy of the Declaration of Conformity is available from Dräger on request.

For Your Safety

The Dräger PAS AirPack 1 Trolley Module incorporates a high performance pressure reducer.  The reducer

is factory set and must not be tampered with, or removed from the manifold, as this invalidates the

conditions of the Dräger warranty.  Correct operational condition is valid only if Dräger service and re-

seals the pressure reducer.

u

Use of this equipment requires wearer training and observance of these Instructions for Use and the

Instructions for Use supplied with the selected associated breathing equipment.

u

Use the equipment for the purpose specified in this instruction, or as confirmed in writing by Dräger.

u

Use and maintenance of this equipment requires knowledge and compliance with National Regulations,

Laws and Standards governing the use of respiratory equipment in the country of use.

u

Only trained competent personnel should inspect and service the equipment at regular intervals and

a record kept of such inspections and service.

u

Only trained and competent personnel should carry out the charging of the compressed air cylinders.

u

Dräger recommends that a Service Contract be obtained from your Dräger Branch or Agent.

u

Contact Dräger for details of Service Contracts and Service Training Courses.

u

Use only original Dräger Spare Parts for service and maintenance.

u

Use only original Dräger Test Equipment for service and maintenance.

u

Notify Dräger if there is a component fault or failure.

Liability Statement

Terms and Conditions of warranty for the Dräger PAS AirPack 1 Trolley Module can be obtained from Dräger

on request.  Responsibility for reliable function of the equipment transfers to the owner or operator when

serviced or repaired by untrained personnel, (not employed or authorised by Dräger), or when used in a manner

not conforming to its intended use.

Description and Intended Use

The Dräger PAS AirPack 1 Trolley Module is a

compact  mobile  two-wheel  trolley  unit  with

flexible rubber airline hose and hose winding drum,

plus storage capacity for two valved high-pressure

cylinder assemblies.  The module provides a low-

pressure independent breathing air supply to the

wearer of approved Dräger breathing apparatus

fitted with airline manifold connections.

The equipment is compatible with a range of

associated  breathing  apparatus  and

accessories and must be used only with air that

complies with the air quality requirements of

respirable breathable air at the pressure and

quantities (volumes) required.  When the airline

fitting of the breathing apparatus is connected

to  the  module,  the  approved  configuration

provides the wearer with respiratory protection

when  working  in  a  contaminated  or  oxygen

deficient gaseous atmosphere.

The manifold assembly of the module allows for

the repeated removal and fitting of cylinders,

transferring of air supply from one cylinder to

another.   This  feature  provides  a  means  of

providing  an  uninterrupted  air  supply  to  the

wearer.

Automatic activation of a preset mechanical high

pressure whistle warning unit (HPWWU) alerts

the Controller of low cylinder pressure.  Activation

of a low pressure whistle warning unit (LPWWU)

alerts the Controller of a drop in pressure of the

low-pressure air supply.

If required, the cylinders may be bypassed by

connecting  the  inlet  coupling  of  the  trolley

module  to  an  independent  low-pressure  air

supply,  e.g.  from  a  works  airline  or  LP

compressor (refer to Main Airline Pressures and

Flow and Air Quality requirement .   A drop in

low  air  supply  pressure  is  highlighted  to the

Controller by the activation of an audible low

pressure whistle warning unit (LPWWU).

The possible available extended working duration,

depends on the configuration selected (attached to

works airline or LP compressor - with filter unit to

maintain air quality), the  capacity (volume) of

cylinder(s) selected, the number of fully charged

cylinders available and the breathing rate of the

wearer(s).

Important Note: No more than two users of

approved  airline  breathing  equipment

should  be  connected  to  the  AirPack  1

Trolley  Module  at  any  time,  and  in  any

configuration.    Refer  to  the  Equipment

Configuration details chart.

Details  of  equipment  variants,  associated

breathing  equipment  and  accessories  are

available from Dräger on request.

Technical Data

High Pressure Connections

Standard G5/8 as per 

EN 144-2

.

200bar

 or 

300bar.

Other  high-pressure  (HP)  connections  are

available to National Standards.

Operating Pressure

Low Pressure - 

8bar 

Nominal. (

6bar to 10bar

)

Whistle Warning Units (WWU)

HPWWU - Preset between 

60bar

 to 

50bar

LPWWU - Preset between 

3bar 

to

 5bar

Compressed Air Cylinders

Contact Dräger for details of approved valved

compressed air cylinder assemblies available for

use with this equipment.

Important Note: Only matched pairs of cylinders

should be fitted to the Dräger AirPack 1

 

Trolley

Module, e.g. 2 x 9 litres 200bar.  Do Not mix

200bar and 300bar cylinders.

Main Airline Pressures and Flow

Safety Note: In order for the Trolley Module to

function efficiently, it is the responsibility of

the  user  to  ensure  that  the  following  air

pressure  and  flows  are  available  from  the

factory mains airline air source.  Air quality

for compressed air breathing systems must

conform to the requirements of EN12021.

The selected LP independent air supply must meet

the following parameters.

u

One User

 – 7bar to 10bar - airflow rate of at

least 550 Litres/minute.

u

Two Users

 – 7bar to 10bar - airflow rate of

at least 550 Litres/minute.

LP Leak Test

 - Indicated pressure should not

decrease more than 

1bar

 in one minute.

Preparation for Use

Only  trained  and  competent  personnel  must

perform the following checks and preparation

procedures before release of the equipment for

operational use to a 'Controller' and potential

breathing apparatus 'Wearer'.

Visual Inspection

Check the cleanliness and integrity of,

u

Trolley frame and wheels.

u

Cylinder support straps (2).

u

Winding Drum and low pressure hose.

u

Pneumatics, Gauges, WWU’s, Quick Release

Coupling(s) (QRC) etc.

u

HP hoses and cylinder connector handwheel

assemblies (2).

Fitting the Cylinders

u

Place the AirPack 1 Trolley Module in the upright

position as shown in Fig. 1.

u

Check that the valve outlet port of each fully

charged  valved  cylinder  is  clean  and

undamaged.  Place each cylinder through the

loop of each of the cylinder straps locating the

rounded end of the cylinders into the cylinder

cradle supports at the base of the trolley frame.

u

Check that both connector handwheels are not

damaged and that the connector O rings are

in position and not damaged.

u

Rotate the cylinders to align the outlet port of

each  valve  with  the  associated  connector

handwheel assembly.  Ensure a smooth curve

of  the  high-pressure  hose.    Screw  each

handwheel into the appropriate cylinder valve

– hand tight.

u

Take up slack of each of the cylinder support

straps then activate the Camlock mechanism

by pulling the free end of the support strap

around the cylinder - securing each cylinder

to the frame.  Fig. 1.

Leak Test and Test of WWU’s

u

'Close'

 both the high-pressure (HP)  bleed

valves  (6)  of  the  cylinder  connector

handwheel assemblies - handtight.

u

'Open'

 slowly, but fully,  

ONE

 of the cylinder

valves to pressurise the system.  Check the

indicated pressure readings on the HP gauge

(2) and the LP gauge (3).

High Pressure

 - Indicated pressure should be

at least 

80%

 of maximum working pressure

indicated on the selected cylinders.

Low Pressure

 - Indicated pressure should be

between 

6bar

 to 

10bar

.

u

'Open'

 the HP bleed valve (6) of the connector

handwheel assembly of the opposite cylinder

(i.e. the ‘Closed’ valve) – there should be no

audible leak.  If satisfactory 

'Close'

 the HP

bleed valve.

u

'Close'

 the cylinder valve.  

Wait for 10 to 15

seconds to allow the system to stabilise then

c

heck the pressure reading on the HP gauge

(2) and the LP gauge (3).

HP Leak Test

 - Indicated pressure should not

decrease more than 

20bar

 in one minute.

u

Following  satisfactory  leak  test,  check  the

activation of the HPWWU (1) and the LPWWU

(4) as follows:

Slowly vent pressure from the HP side of the

system by gripping and carefully pulling the

handle of the relief valve (7).  As the pressure

decreases observe the HP gauge (2) .

HPWWU

 - the whistle should begin to sound

between 

60bar

 to 

50bar

.  As soon as the whistle

sounds, release grip on the handle of the relief

valve (7) and allow the whistle to vent.  When

the  HPWWU  stops  sounding    proceed

immediately to test the LPWWU.

u

Following satisfactory HPWWU test, check the

activation of the LPWWU as follows:

Again grip and carefully pull the handle of the

relief valve (7).  As the pressure decreases

observe the LP gauge (3) .

LPWWU

 - the whistle should begin to sound

between 

5bar

 to 

3bar

.

u

Check the HP bleed valves (6) at each cylinder

valve are 

'Closed'  

then slowly, but fully, 

'Open'

the 

second

 cylinder valve to pressurise the

system.  Check the indicated pressure readings

on the HP gauge (2) and the LP gauge (3).

High Pressure

 - Indicated pressure should be

at least 

80%

 of maximum working pressure

indicated on the selected cylinders.

Low Pressure

 - Indicated pressure should be

between 

6bar

 to 

10bar

.

u

Open the high-pressure bleed valve (6) of the

connector handwheel assembly of the opposite

(first) cylinder (i.e. the ‘closed valve.) – there

should be no audible leak.  If satisfactory 

'Close'

the bleed valve.

u

'Close'

 the cylinder valve and then check the

pressure reading on the HP  gauge (2).

High Pressure Leak Test

 - Indicated pressure

should not decrease more than 

20bar

 in one

minute.

u

Following satisfactory completed tests, grip and

pull the handle of the relief valve (7) to vent

pressure from system.

u

Having ‘passed’ the tests, the equipment is now

available for use.

Pre-Use Procedures

Caution:  Before connecting the selected airline

breathing equipment to the tested and released

Dräger 

PAS AirPack 1

 Trolley Module, ‘check’

that the breathing equipment has been tested

by  a  trained  and  competent  person  before

being  released  for  use.    When  in  use  the

AirPack 1

 

Trolley Module must be positioned

in a safe and uncontaminated area.  Dräger

recommend that a ‘Controller’ must be present

at all times to monitor and maintain the air

supply to the wearer of the selected breathing

equipment.  Time required for the wearer of

the  airline  breathing  equipment  to  safely

evacuate from the hazardous area to a safe area

must be within the air capacity of the remaining

cylinder(s) of the 

PAS AirPack 1

 Trolley Module.

Dräger recommend that the user(s) of this final

configuration operate and maintain a ‘Permit

for  Use’  procedure  and  also  have  an

established company evacuation procedure(s).

The Dräger PAS AirPack 1 Trolley Module may be

used either as a self-contained unit, i.e. using the

attached compressed air cylinders, or by connecting

the module to an independent low-pressure air supply,

e.g. from a works airline or LP compressor the

cylinders can be bypassed.

Self-contained Unit

The Controller

u

With the trolley module suitably positioned in a

safe and uncontaminated area, pull and unwind

a required length of hose from the winding drum.

Attach any required accessories, e.g. extension

hoses, 'Y' pieces, etc.

u

Check that both HP bleed valves (6) at each of

the  connector  handwheel  assemblies  are

'Closed'

 (clockwise).

When used as Self-contained Unit

u

'Open'

 slowly, but fully,  

ONE

 of the cylinder

valves to pressurise the system.  Check the

indicated pressure readings on the HP gauge

(2) and the LP gauge (3).

High Pressure

 - Indicated pressure should be

at least 

80%

 of maximum working pressure

indicated on the selected cylinders.

Low Pressure

 - Indicated pressure should be

between 

6bar

 to 

10bar

.

When used from a works airline or

LP compressor

Safety Note: Refer to Airline Pressure and

Flow.    Air  quality  must  conform  to  the

requirements of EN12021.  If required use the

Dräger Filter Unit.

u

Connect an approved extension hose from the

selected LP independent air supply to the male

connector (5) of the trolley manifold.  

'Open'

the valve of the independent air supply to

pressurise  the  system.    Check  indicated

pressure reading on the LP gauge (3).

Low Pressure

 - Indicated pressure should be

between 

6bar

 to 

10bar

.

The Controller

Important Note: Should the 'Controller' be

required  to  wear  respiratory  protection

equipment  then  all  the  following  'Wearer'

instructions should be observed.  Connect the

male  coupling  of  the  selected  breathing

equipment to the female QRC (8) of the pressure

manifold.  This will reduce the duration available

from the selected system.   It must be observed

however  that  no  more  than  two  users  of

approved airline breathing equipment should

be connected to the PAS  AirPack 1 Trolley

Module at any time, and in any configuration.

The Wearer

u

Refer to the Instruction for Use for the selected

breathing equipment, put on the equipment and

prepare the equipment for attachment and use

with the PAS AirPack 1 Trolley Module.

u

Connect the male coupling of the breathing

equipment to the female QRC of the supply

hose from the trolley module, i.e. the winding

drum.

Safety Warning: Facial hair, beard stubble,

side-whiskers, and the wearing of spectacles

will  adversely  affect  and  interfere  with

facepiece seal. Correct fit of facepiece only

ensured if facepiece seal makes close contact

with skin.

u

Put on the facepiece as defined in the Instruction

for Use for the selected breathing equipment,

breathe normally and perform Function Check.

Function Check

The Controller

u

'Close'

 the cylinder valve or the valve of the

associated air supply.

The Wearer

u

Breathe normally to vent air from the system.

During venting the whistle(s) will alarm at the

preset  pressure.    When  fully  vented,  hold

breath.  The facepiece should hold onto the face

indicating a positive seal.

u

Immediately indicate to the ‘Controller’ of a

positive face seal and request to immediately

‘Open’

 the valve of the associated air supply.

The Controller

u

When ‘Wearer’ indicates to the ‘Controller’ of a

positive face seal – the ‘Controller’ should

immediately 

‘Open’

 the associated valve to

pressurise the system.

The Wearer – In Use

u

Inhale and hold breath.  The unit must balance,

i.e. no audible leak.  Continue to breathe normally.

Expired air should flow easily out of the exhalation

valve of the facepiece.  Press the centre of the

rubber  cover  of  the  LDV  to  check  for

supplementary supply airflow then release.

u

Continue to breathe normally and as instructed

by the controller proceed to the working area

taking care to support and route the airline hose

as necessary.

In Use - When used as a Self-

contained Unit

The Controller

u

Regularly take a HP gauge reading.  At between

60bar

  to 

50bar 

the  HPWWU  unit  sounds

indicating that the cylinder in use is almost

empty.  To therefore maintain the air supply to

the  wearer  of  the  breathing  equipment

immediately proceed as follows:

1

. Slowly, but fully, 

'Open'

 the valve of the

second

 fully charged cylinder.  The HP gauge

(3) will indicate an increase in system pressure.

The  wearer  is  now  connected  to  the  fully

charged cylinder.

2

'Close' 

the valve of the empty cylinder then

'Open'

 the bleed valve (6) of the associated

valve to vent the remaining pressure from the

HP supply hose.

3. Unscrew the connector handwheel from the

valve of the depleted cylinder.  Release the

Camlock mechanism of the cylinder support

strap and remove the cylinder.

4. Select an available replacement fully charged

cylinder and assemble to the trolley.  See -

Preparation for Use – Fitting Cylinders.

5. 

'Close'

 the bleed valve (6) of the connector

handwheel  assembly  of  the  replacement

cylinder.

Note: 

To maintain a supply of respirable breathing

air to the wearer of the breathing equipment, this

procedure can be repeated with further available

replacement fully charged cylinders.

Safety Warning: Should the LPWWU of the PAS

AirPack  1  Trolley  Module  sound,  then  the

controller must immediately implement the

agreed and established company evacuation

procedure(s).

The Wearer – Self Contained Operation

u

Breathe normally.

u

When the task is completed, or when advised

by the ‘Controller’, proceed to the safe area

(Controller) by the safest route.

Important  Note:  -  Refer  to  the  evacuation

procedures as defined in the Instructions for

Use for the selected breathing equipment.

In  Use  -  When  used  from  a

Works Airline or LP Compressor

Safety Note: Refer to Airline Pressure and Flow.

Air quality must conform to the requirements

of EN12021.  If required use the Dräger Filter

Unit.

The Controller

u

Take regular LP gauge (3) reading.  The indicated

pressure should be between 

6bar

 to 

10bar

.

Safety Warning: Should the LPWWU sound,

indicating a reduction in pressure from the

Works Airline or LP Compressor, the ‘Controller’

must    immediately  fully  'Open'  ONE  of  the

cylinder valves of the 

PAS AirPack 1

 Trolley

Module to pressurise the system.  The extension

hose from the selected LP independent air

supply to the male connector (5) should be

disconnected.

u

Continue to follow the instructions outlined in

the section – i.e. 

In Use- When used as Self-

contained Unit.

The Wearer -  Works Airline etc.

u

Breathe normally.

u

When the task is completed, or when advised

by the ‘Controller’, proceed to the safe area

(Controller) by the safest route.

Important  Note:  -  If  necessary  refer  to  the

evacuation  procedures  as  defined  in  the

Instructions for Use for the selected breathing

equipment.

After Use

The Wearer

u

Refer  to  the After  Use  instructions  in  the

appropriate Instruction for Use for the selected

breathing equipment.  Remove the equipment

and detach the male coupling of the breathing

equipment from the female QRC of the supply

hose of the carrying frame module.

Safety Warning:  Do Not attempt to remove the

breathing equipment until in safe area, clear of

the hazard.

The Controller

u

Following  the  removal  of  the  breathing

equipment by the wearer either:

1.

disconnect  the  extension  hose  from  the

selected LP independent air supply from the

male connector (5).

or

2.

 

'Close'

 the relevant cylinder valve of the

trolley module.

u

 Grip and pull the handle of the relief valve (7) to

vent pressure completely from system.  

'Open'

the (HP)    bleed  valves  (6)  of  the  cylinder

connector handwheel assemblies

u

Carefully and neatly rewind the airline hose

to the winding drum.  Dräger recommend that

any dirt or contaminants on the hose should

be removed as it is being rewound onto the

winding drum.

Routine Maintenance

To be performed after use of the equipment.

See also Maintenance and Test Intervals Chart.

Refer to the Instructions for Use supplied with the

selected breathing equipment, facepiece, lung

demand valve etc.

Visual Inspection

Check the cleanliness and integrity of,

u

Trolley frame and wheels.

u

Cylinder support straps (2).

u

Winding Drum and LP hose.

u

Pneumatics, Gauges, WWU’s, Quick Release

Coupling(s) (QCR) etc.

u

HP hoses and cylinder connector handwheel

assemblies (2).

u

Associated extension hoses.

Cleaning, Disinfecting, Drying

Dräger recommend that contaminated and dirty

components  or  assemblies,  hoses  etc.  are

carefully cleaned and disinfected then thoroughly

dried after use, or as considered necessary by the

user.

To ensure correct operational condition of the

equipment  use  the  cleaning  and  disinfecting

solutions recommended by Dräger.  The use of

any  other  product  will  invalidate  the  Dräger

warranty and guarantee.

Safety Note: Refer to manufacturers’ usage

instructions  when  using  cleaning  and

disinfecting  agents.  It  is  important  that

attention be paid to concentration and reaction

times. Do Not use organic solvents, such as

Acetone, 

Alcohol, 

White 

Spirit,

Trichloroethylene or similar.

Dräger recommends the following:

1. Cleaning

u

Dräger Safety Wash.

u

Sekusept

Do Not

 exceed a temperature of 

30 degrees

Celsius.

Note:

 

Before  disinfecting,  rinse  off  cleaning

solution in clean water.

2. Disinfecting

u

Incidur

Do Not

 exceed a temperature of 

30 degrees

Celsius.

Note:

 

Before drying, rinse off 

disinfecting

 solution

in clean water.

u

Dräger Wipex Cloths.

Details of cleaning and disinfecting agents are

available from Dräger on request.

3. Rinsing and Drying

Remove cleaning and disinfecting solutions by

rinsing in clean water, followed by drying.

Do Not

 exceed a temperature of 

60  degrees

Celsius when drying components.

Charging Cylinders

Refer to 'For Your Safety'

Safety Warning: Air quality for compressed air

breathing  systems  must  conform  to  the

requirements of EN12021.

Only charge compressed air cylinders which:

u

Conform to National Standards.

u

Feature the original manufacturers test date

and test mark.

u

Have not exceeded the test date indicated

on the cylinder by the last testing station.

u

Do Not used damaged cylinders.

Recharge to the rated pressure indicated on the

label  or  stamped  on  the  neck  or  shoulder  of

cylinder.  Dräger recommend a charge rate of

27bar/minute.    Rapid  charging  will  induce  an

increase in temperature resulting in an incomplete

charge - check pressure at ambient and if required

‘Top Up’ charge.

To prevent overcharging of the selected cylinder

Dräger recommend that a pressure-limiting device

is fitted to the charging compressor.

Storage - Ready for Use

u

Store the Trolley Module in an upright position

as shown in Fig. 1.  A  protective dust cover is

available.  Contact Dräger for details.

u

Store the equipment ‘Ready for Use’ in a cool

dry  environment,  free  from  dust  and  dirt.

Protect rubber parts - 

Do Not

 expose to direct

sunlight.

7

5

6

3

2

4

1

8

Reviews: