Dräger P7-TD – Quick Reference Guide

 

 

Any use of the Dräger P7-TD requires the full understanding of and compliance with the 
instruction manual 5695075 (P7-TD) for this device.

 

The instruction manual can be found on our website 

www.draeger-msi.de/en

 in the menu item 

Services

 

>Downloads>P7 >Operating Instructions. 

You can find the measurement data management software PC200P, which you can download after a brief 
registration using your device number and your address data, in the menu item 

Services >Downloads >P7> 

Software

.

 

 
 

Only charge the Dräger P7 using the corresponding charger. 

 

Key functions 

 
Switching on:

 Simultaneously press the keys “F” and “H” for 1 second. 

Switching off:

 Press the “F” key for more than 3 seconds or select “Switch off” in the “Functions” area. 

 

All other key functions are shown on the display directly using the keys!

 

 

Functions 

The individual tests and measurements are accessed via the “Functions” menu. 
 

Pressure measurements 

 

1.  High pressure up to 0.12 MPa (1.2 bar) 

Use pressure hose with 

pneumatic quick coupling NW5

 

 

2.  Medium pressure up to 150 hPa (mbar) 

 

3.  Micro pressure up to 100 Pascal 

 

4.  Medium pressure up to 150 hPa (mbar) with pump function 

 

5.  High pressure up to 0.35 MPa (3.5 bar) or 2.5 MPa (25 bar) 

(with external high pressure sensor) 

 

Follow the instructions for the Dräger P7 for pipeline testing! 

 

Manual tightness test

 

with absolute pressure and temperature compensation

 

 

Test pressure, stabilization period and measurement period can be set during the manual tightness test. 
 
1. Select measuring channel 0 - 150 hPa (mbar), 0 - 3,500 hPa (mbar) or 0 - 25,000 hPa (mbar) 
2. Select test pressure: 10 - 150 hPa (mbar), 100 - 3,500 hPa (mbar) or 1,000 - 25,000 hPa (mbar) 

If a tightness test was selected in the 0 - 150 hPa (mbar) measuring channel, the internal or an external 
pump can be used, to build up the test pressure. Tests with the 3,500 hPa (mbar) or 25,000 hPa (mbar) 
channels can only be conducted using an external pump.  

3. Select stabilization period: 00:01 - 12:00 h 
4. Select measurement period: 00:01 - 12:00 h 