Maintenance
6. Enter the password (password when delivered = 001).
7. Using the
key, select the standard sensitivity calibration
function »
«.
The symbol for standard sensitivity calibration flashes.
8. Press the
key to start the sensitivity calibration.
The name of the gas flashes.
9. Confirm with the
key.
The calibration gas concentration is displayed
10. Press the
key to confirm the calibration gas concentration or
use the
key to change the calibration gas concentration and
complete the process by pressing the
key.
The set calibration gas concentration flashes.
11. Press the
key to confirm the set value.
12. Open the valve on the test gas source to let test gas flow over
the sensor.
The currently displayed measured values start to flash.
The flashing stops after a static measured value has been
reached.
The displayed measured values change to the values
according to the gas supplied.
When the calibration is complete and the displayed measured
values have stabilized:
The display of the current gas concentration alternates with the
» OK «
display.
13. Press the
key or wait for 5 seconds to quit the calibration.
The instrument changes to the measuring mode
14. Close the valve on the test gas source.
If a fault has occurred during the sensitivity calibration:
The fault message » « appears and »
« is displayed for the
respective sensor instead of the measured value.
Repeat the sensitivity calibration.
If necessary replace the sensor; see the Technical Handbook.
6
Maintenance
6.1
Maintenance intervals
The instrument should be inspected and maintained by suitably
qualified persons annually (consult: EN 60079-29-2 – Guide for the
selection, installation, use and maintenance of apparatus for the
detection and measurement of combustible gases or oxygen;
EN 45544-4 – Electrical apparatus used for the direct detection and
direct concentration measurement of toxic gases and vapours - Part 4:
Guide for selection, installation, use and maintenance and national
regulations).
Calibration intervals: see Instructions for Use for the respective
DrägerSensors.
6.2
Cleaning
The instrument does not need any special care.
If the instrument is very dirty, clean it with a cloth.
OK
OK
OK
OK
OK
OK
CAUTION
Abrasive cleaning implements (brushes etc.), cleaning agents
and cleaning solvents can destroy the dust and water filters.
!
Summary of Contents for MQG 0020
Page 2: ......
Page 28: ...28 Dr ger X am 5100 Technical Data...
Page 69: ...Dr ger X am 5100 69 Dados t cnicos...
Page 158: ...158 Dr ger X am 5100 1 6 170 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger 1 1 i i...
Page 163: ...Dr ger X am 5100 163 3 1 5 1 2 Bump Test 21 O2...
Page 166: ...166 Dr ger X am 5100 166 3 6 3 6 1 A1 A1 A1 1 2 3 6 2 A2 A1 i i OK...
Page 170: ...170 Dr ger X am 5100 6 6 1 EN 60079 29 2 EN 45544 4 4 Dr gerSensor 6 2...
Page 220: ...220 Dr ger X am 5100 1 6 232 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger 1 1 i i...
Page 225: ...Dr ger X am 5100 225 3 1 5 1 2 Bump Test 21 O2...
Page 228: ...228 Dr ger X am 5100 9 A1 3 5 228 3 6 3 6 1 A1 A1 A1 A1 A2 3 6 2 A2 A1 i i OK...
Page 232: ...232 Dr ger X am 5100 6 6 1 EN 60079 29 2 EN 45544 4 4 Dr ger 6 2 7 Dr ger Dr ger...
Page 259: ...Dr ger X am 5100 259 1 6 271 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger 1 1 i i...
Page 264: ...264 Dr ger X am 5100 3 1 4 3 2 1 3 1 5 1 2 Bump Test Bump Test OK 21 Vol O2...
Page 270: ...270 Dr ger X am 5100 5 5 6 001 7 8 9 10 11 12 OK 13 5 14 OK OK OK OK OK OK...
Page 271: ...Dr ger X am 5100 271 6 6 1 EN 60079 29 2 EN 45544 4 4 Dr gerSensor 6 2...
Page 286: ...286 Dr ger X am 5100 Notes on Approval A 2 Declaration of Conformity...
Page 287: ......