For your safety and that of your patients
Electromagnetic compatibility (EMC)
Medical electrical equipment is subject to special
precautionary measures concerning
electromagnetic compatibility. Observe the EMC
information during installation and initial operation.
For further information, see the following section:
"EMC declaration" (page 40).
Mobile high-frequency communication equipment
can affect medical electrical equipment.
Installing accessories
Strictly observe instructions for use and assembly
instructions.
Training
Training for users is available from the Dräger
organization responsible, see www.draeger.com.
Installation information
Contact information for assembly personnel is
available from DrägerService.
The assembly instructions are available to the
organization operating the healthcare facility.
Product-specific safety information
WARNING
Risk of device malfunction
Install accessories to the basic device in
accordance with the instructions for use of
the basic device. Make sure that there is a safe
connection to the basic device.
WARNING
Components that have been subsequently
installed on the medical device must comply
with all regulatory or technical requirements
and certifications and are the responsibility of
the organization operating the healthcare
facility.
Dräger assumes no warranty and no liability
for parts from third-party manufacturers or the
functionality of third-party manufacturers'
products.
Only experts may carry out subsequent
installation.
If the above are not complied with, the correct
functioning of the medical device may be
compromised.
WARNING
Risk of electric shock
The device housing must not be opened.
WARNING
Increased fire risk
Gas terminal units must not come into contact
with oil, grease or flammable liquids.
WARNING
Danger of overheating
The support arm can overheat if objects are
placed on the support arm in the area of the
light fittings.
Do not place objects on the support arm.