Assembly and preparation
16
Instructions for use Movita
Assembly and preparation
Positioning and connecting devices
Place the additional devices on the rack below
the media head or on the media column. Do not
exceed the maximum load in doing so.
Position the devices symmetrically on the rack.
The load on the rack must be even.
Connecting the devices to an electricity
supply
Device combinations must comply with the
requirements of these instructions for use, see
"Connection to other devices" on page 6.
WARNING
Danger of damage to devices or personal
injury due to the ceiling supply unit rotating
on its own.
If extreme changes to the load situation of the
ceiling supply unit occur during operation
(pertains only to ceiling supply units equipped
exclusively with a friction brake), the ceiling
supply unit may rotate on its own.
After any extreme change to the load
situation, the arm systems of the ceiling
supply unit must be readjusted by experts
according to the assembly instructions.
WARNING
Danger of overloading the ceiling supply unit
Do not exceed the maximum load for the
supply unit and individual devices when
positioning the intended devices.
The maximum load is based on the customer-
specific assembly of the supply unit.
CAUTION
If the rack is loaded asymmetrically, the rack can
rotate on its own.
WARNING
Risk of device malfunction
If a connected device trips the on-site
automatic circuit breaker of the respective
circuit, all other connected devices in the
same circuit will also no longer be supplied
with current.