8
Supplement to the instructions for use
|
Isolette 8000 plus
Reprocessing
3. Remove the mattress (B).
4. If the incubator is equipped with a weighing scale, perform steps 5 through 7;
otherwise go to step 7.
5. Disconnect the scale cable from the sensor module.
6. Remove the cable from the clamps that hold it to the incubator.
7. Lift the scale (C) from the bed.
8. Remove the bed (D) and X-ray tray (inside bed)
9. Remove the T-bars (E).
10. Remove upper cover (F).
Removing the heater radiator and impeller
1. On devices without the humidification system option, remove the heater/impeller
cover (A).
2. On devices with the optional humidification system, pull the water reservoir
partially out, and remove the heater/impeller cover (B) and the duct cover (C).
WARNING
Risk of death or serious injury
The heater can be sufficiently hot to cause burns.
►
Avoid removing or touching the heater until the device has been switched off
for at least 45 minutes.
3. When the device has cooled, remove the heater radiator (D).
4. Pull the impeller (E) off the motor shaft.
48107
A
B
C
D
E