22
Supplement to the instructions for use
|
Isolette 8000 plus
Reprocessing
Fitting the heater radiator and fan impeller
1. On devices without the humidification system option, install the impeller (A) onto
the motor shaft.
2. Install the heater radiator (B).
3. On devices without the humidification system option, install the heater/impeller
cover (C).
4. On devices with the humidification system option, install the heater/impeller
cover (D) and the duct cover (E).
WARNING
Risk of death or serious injury
Using older software versions can lead to incorrect operation of the humidification
system leading to injury.
►
The humidification system with the heater/impeller cover with duct cover must
be used with software version 5.n or higher.
Fitting the scale and the bed and related components
NOTICE
Risk of damage to the device
The hood contains 4 tabs (one at each corner) which keep the deck securely in
place. Installing the upper cover when the hood is down may result in damage to
the upper cover and/or jamming of the hood in place.
►
Ensure that the hood is raised before attempting to install the upper cover.
48252
C
D
E
B
A