Supplement to the instructions for use
|
Isolette 8000 plus
19
Reprocessing
Dräger states that oxygen-releasing agents and chlorine-releasing agents may
cause color change in some materials. Color change does not indicate that the
product is not functioning correctly.
Other surface disinfectants are used at one's own risk.
For disinfectant recommendations for the SoftBed mattress, consult their
manufacturer instructions for use.
2.8 After
reprocessing
2.8.1
Assembling and fitting device-specific components
Prerequisites:
– All components have been reprocessed and dried.
– All components have been inspected for breakage or cracks, and replaced if
necessary.
Assembling the water reservoir assembly
1. On devices with the humidification system option:
a. Install the humidifier (A) into the water reservoir assembly. Insert it fully into
the humidifier chamber and secure it by closing the 2 latches (B).
b. Install the lid on the water reservoir and push back the latch (C) on top of the
lid to secure it.
c. Insert the water reservoir assembly in the shell with the latch (D) open. Then
slide it in fully and lock it in place by pushing in the latch.
49650
48251
A
B
B
C
D