12
Supplement to the instructions for use
|
Isolette 8000 plus
Reprocessing
Removing the sleeves (if installed), hand port gaskets, and hose grommets
1. Open the hand port doors and remove the sleeves, if installed.
2. Pull the hand port gaskets (A) from all sides of the hood to remove them.
3. Pull the hose grommets (B) from each side of the hood to remove them.
4. Wipe the retainer rings clean, if installed.
5. Discard disposable sleeves, if installed.
Removing the air inlet filter
1. Loosen the two thumbscrews (A) on the air inlet filter cover located under the
shell assembly on the rear of the device.
2. Remove the air inlet filter cover.
3. Remove the air inlet filter.
4. Discard the air inlet filter.
49649
48102
A
B
A
A