4-4
D
ELTA
/D
ELTA
XL/K
APPA
VF8
4 A
DMISSION
, T
RANSFER
,
AND
D
ISCHARGE
Data Transfer Using the Memory Card
Transferring data from one monitor to another with the PCMCIA memory card is a
two-step process:
1. Copy data from the source monitor to the card.
2. Copy the data from the card to the destination monitor.
After the data has been copied to the destination monitor, it is no longer available on
the card. The monitor displays the current patient’s name and ID number at the
beginning of a data transfer. Because the data on the card overwrites data on the
destination monitor, you can overwrite one patient’s data with another’s, effectively
discharging the former and admitting the latter. Make sure you copy information to
the destination monitor before you perform significant monitoring functions.
Memory Card Transfer
WARNING:
Use electrostatic discharge (ESD) prevention practices
when inserting the PCMCIA card into the monitor. In
some environmental conditions, insertion of the
memory card could cause the monitor to reset as the
result of an ESD event.
The patient’s stored event and trend information will be
lost after the monitor resets.
Monitoring does not occur during data transfer.
CAUTION:
Do not remove the memory card while a copy is in process. If the transfer
fails, repeat the procedure using a new card.
NOTE:
A memory card transfer from a monitor with VF8 level software to a monitor
with a lower level software functions appropriately. However, a memory card transfer
from a monitor with a lower level software to one with a VF8 level software is no
supported.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...