Connecting to the Network
Connecting the monitor to the network via the Infinity Docking Station (IDS) gives
you access to the following:
Power
Infinity network
Bedside recorder
Nurse call alarm
Remote keypad
Memory for storing monitor setup defaults
Scio MultiGas modules for anesthetic and respiratory monitoring
MIB device interfaces
The DirectNet feature allows you to connect your monitor directly to the Infinity
network, bypassing the need for a Docking Station or Infinity Docking Station.
DirectNet does not support the MultiGas module or MIB protocol.
Connecting the Delta/Delta XL to the Network
For the Delta/Delta XL, you can also use a Docking Station
™
, to access the network
(refer to the
Infinity Docking Station
and
Docking Station
Power Supply Hardware
Installation Instructions).
Connecting the monitor to the
network:
1. Place the monitor on the IDS or Docking
Station using both hands — one holding
the handle, the other steadying the
monitor. Make sure the monitor clicks
firmly into place.
2. Slide the lever to the right to lock the
monitor in place.
Be sure the monitor is securely positioned, as the
lever does not move unless the monitor is seated properly. A battery charge indicator
LED lights up when the monitor is properly docked.
IDS
or
Docking
Station
Lever
Move right to
lock
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...