Quick Reference Table – OR Menu
OR Function
Description
Settings
The OR menu
OR
Activates the OR menu functions.
NOTE
: This function is a locked
option. It is installed at the time of
purchase from perioperative care unit
or it can be installed locally by
DrägerService or your hospital’s
technical personnel.
•
ON
– OR functions are enabled.
•
OFF
– The monitor reverts to normal
functions; The settings
Cardiac
Bypass
and
NBP Chime
are
ghosted.
Cardiac Bypass
Configures the monitor for use during
cardiac surgery.
•
ON
– Suspends all patient-
monitoring alarms (except ventilator
alarms), NBP interval
measurements, and arrhythmia
detection.
•
OFF
– The monitor reverts to normal
functions.
NBP Chime
Controls whether an attention tone
sounds upon completion of an NBP
measurement, see chapter 13 for
more information).
•
ON
– Attention tone sounds when
NBP measurement is complete.
•
OFF
– No tone sounds when NBP
measurement is complete.
Alarm Volume
Sets the alarm volume.
•
10 - 100%
in increments of 10
• OFF
Large IBP-Mean
Display
Determines the relative size of the
mean pressure value in invasive
pressure parameter boxes.
•
ON
– The mean IBP value is larger
than the systolic and diastolic IBP
values.
•
OFF
– The systolic, diastolic and
mean IBP values are the same size.
Attention Tone
Volume
Sets the attention tone volume
•
Off, 5, 10 - 100%
in increments of 10
HR Source
Derives heart rate from various
sources (see page 8-23).
NOTE
: This function is useful during
electrosurgery when artifact makes
the ECG channel unavailable.
Click on one of the following settings to
determine the Heart Rate source:
• ECG
• ART
• SpO2
• AUTO
Filter
Determines sensitivity to noise,
artifact, and other signal distortion
(see page 8-20).
NOTE
: The
ESU
setting automatically
disables pacer detection.
• OFF
• ESU
• Monitor
ARR Monitoring
Determines the number of arrhythmia
events you can monitor
(see page 9-5).
• OFF
• Basic
• Full
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...