![Dräger Infinity Delta Instructions For Use Manual Download Page 548](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779548.webp)
B-22
D
ELTA
/D
ELTA
XL/K
APPA
VF8
B T
ECHNICAL
D
ATA
etCO
2
Pod
Physical Attributes
Size (H x W x D):
Pod: approx. 140 x 140 x 51 mm (5.5 x 5.5 x 2 in)
Capnostat
TM
III Sensor: 33 x 42 x 22 mm (1.3 x 1.7 x 0.9 in)
Weight:
Pod:
0.5 kg (1.1 lb)
Capnostat
TM
III sensor: 18 g
Connections:
Sensor connector, female luer side-stream sampling port, male
luer sample exhaust port
Adult airway adapter dead
space:
< 5 cc
Neonatal airway adapter dead
space:
< 0.5 cc
Moisture resistance:
Airway adapter can be immersed in water without damage
Biotoxicity:
None per ASTM F720 (MTL #2009)
NOTE:
Capnostat
TM
III sensor size and weight exclude cable.
Electrical Specifications
Power source:
Powered directly from monitor
Protection against electric
shock:
Type CF
Mode of operation:
Continuous
Defibrillation protection:
Defibrillation-proof applied part per IEC 60601-1
Environmental Requirements
Temperature range:
Operating: 10 to 40 °C (50 to 104 °F)
Storage:
–20 to 50 °C (–4 to 122 °F)
Relative humidity:
Operating: 20 to 90 %, non-condensing
Storage:
10 to 95 % (with packaging)
Atm. pressure:
Operating: 525 to 795 mmHg (70 to 106 kPa)
Storage:
375 to 795 mmHg (50 to 106 kPa)
NOTE:
Reading complies to BTPS = body temperature 37 °C, ambient barometric pressure of
750 mmHg, and relative humidity of 100 %.
Regulatory Requirements
Standards:
IEC 60601-1, ISO 21647:2004 with Corri. 1:2005 (applicable
sections only, Oxygen analyzers for monitoring patient breathing
mixtures); EN ISO 21647:2009
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...