![Dräger Infinity Delta Instructions For Use Manual Download Page 529](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779529.webp)
O
VERVIEW
VF8
D
ELTA
/D
ELTA
XL/K
APPA
B-3
Overview
This appendix contains technical specifications for the physical and functional aspects
of the patient monitoring system. These specifications apply to adult, pediatric, and
neonatal patients.
Upon request, Dräger makes any technical information required to perform
maintenance and/or calibration of serviceable items available to qualified technical
personnel.
Overall Regulatory Standard Compliance
IEC 60601-1 and applicable collateral and particular standards
IEC 60601-1-2
Electromagnetic Compatibility (EMC)
This section is intended to provide information with regard to electromagnetic
compatibility for the Dräger Delta series of patent monitors (hereafter referred to as
'equipment'). It supplements the information that already exists elsewhere in the
instructions for use.
Much of the information below is derived from requirements specified in the
electromagnetic compatibility standard for medical electrical equipment IEC 60601-1-
2: 2001/A1:2004 published by the International Electrotechnical Commission and,
available from a variety of sources. While primarily aimed at device manufacturers
this contains a large amount of information that may be useful to interested users of
medical equipment.
The information contained in this section (such as separation distances) is in general
specifically written with regard to the Dräger patient monitors specified above. The
numbers provided will not guarantee, but should provide reasonable assurance of,
faultless operation. This information may not be applicable to other medical electrical
equipment, older equipment may be particularly susceptible to interference.
NOTE:
Medical electrical equipment needs special precautions regarding
electromagnetic compatibility (EMC) and must be installed and put into service
according to the EMC information provided in this section and the instructions
for use which accompanied your monitor.
Portable and mobile RF communications equipment can affect medical
electrical equipment.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...