Independent Surgical Display (ISD)
The Independent Surgical Display (ISD) is a special interface that allows patient
monitoring data to be shown on a remote video display. The Surgical Display
Controller connects the patient monitor to the remote video display, and allows you to
access and use the ISD interface.
A non-medical grade video display may be used if
that display and the Surgical Display Controller are
properly positioned. The video display must comply
with IEC 60950 rather than IEC 60601-1 and the
Surgical Display Controller must be kept out of the
patient's vicinity.
To access the Surgical Display main
menu
1. Press the
Menu
fixed key on the front of the
bedside monitor.
2. Click on
Surgical Display
.
The configuration of ISD menus does not affect other monitor functions, displays,
menus, or setups.
SEF
Spectral Edge Frequency
Hz
Power
Total Signal Power
dB
NOTES
:
• SEF and Power are available for viewing only on the Show All Parameters display.
• EEG parameters do not alarm.
NOTE:
The ISD only displays those parameters which appear on the ISD Parameter
Priority screen. Set ISD parameter priority using the procedures described on
page 2-5.
The ISD does not display units of measure, alarm status (alarm limits or alarm
off icons) or visual indications that alarms are occurring.
There is a short (200 ms) but perceptible time delay between the waveforms
shown on the bedside monitor and the corresponding waveforms on the
Surgical Display.
Label
Parameters Supported
Units
To remote video display
To IDS
Surgical Display Controller
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...