Quick Reference Table – BIS Setup menu
Menu Item
Description
Available Settings
Show all
parameters
Opens a separate menu with all BISx
parameters displayed.
BIS:
Bispectral Index
(range = 0 - 100)
continuous processed EEG
parameter that correlates to
a patient’s level of
consciousness, where
100=awake and 0=flatline
EEG.
SQI:
Signal Quality Index
(Range = 0 - 100 %)
a measure of the signal
quality for the EEG channel
source, which is calculated
based on impedance data,
artifact, and other variables.
It is displayed in the form of
a simple bar graph, located
at the bottom of the
Parameter box, above EMG.
EMG:
Electromyograph Indicator
(Range = 0 - 100 dB)
The power (in decibels) in
the frequency range 70-
110Hz is labeled “EMG”.
This frequency range
contains power from muscle
activity as well as power
from other high-frequency
artifacts. It is displayed as a
simple bar graph, located at
the bottom of the parameter
box, below SQI.
BSR:
Suppression ratio
(range = 0 - 100 %)
A calculated parameter to
give the user an indication
when an isoelectric (flatline)
condition may exist.
Suppression ratio is the
percentage of time over the
last 63-second period that
the signal is considered to
be in a suppressed state.
For example BSR=11
(isoelectric over 11% of the
last 63-second review, or 7
seconds).
SEF:
Spectral Edge Frequency
(range = 0.5 - 30.00 Hz)
The frequency at which 95%
of the total power lies below
it and 5 % lies above it.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...