Quick Reference Table -- NMT Monitoring Settings
Menu Item
Description
Available Settings
Start NMT
Starts NMT
measurements.
Not applicable
Stop NMT
Stops NMT measurements. Not applicable
Mode
Allows you to select
measurement mode.
• Single
•
TOF
(default)
• PTC
Measurement Interval
Allows you to select time
interval for automatic NMT
measurement cycles (in
single and TOF modes).
None, 1s, 10s, 20s (default), 1min,
5min, 15min, 30min
NOTE
: in PTC mode, you cannot select
a measurement interval; in TOF mode,
1s is not available.
Pulse Width *
Allows you to select
duration for stimulation
pulse.
100 µs, 200 µs
(default)
, 300 µs
Stim Current Mode
Allows you to select
stimulation current mode.
• In auto mode, the pod
establishes a
supramaximal current
during the first
measurement and uses it
for subsequent
measurements.
• In manual mode, the user
selects a stimulation
current.
•
Auto
(default)
• Manual
Stim Current *
Allows you to select
stimulation current
(read-only field in Auto
mode)
•
Auto mode
: none
(displays the stimulation current value)
•
Manual mode
:
5 mA
(default) -
60 mA
,
(in increments of 5 mA)
*
Caution
: Changing the stimulation current or pulse width invalidates the stored reference value.
The pod recalculates the reference twitch at the next measurement, based on the existing level of
patient relaxation, which could cause misleading NMT measurements.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...