To start NMT measurements
Click on
Start NMT
.
The pod automatically establishes a supramaximal current and reference twitch (see
page 24-2). Once these have been found and stored, NMT measurements in the TOF
mode begin automatically at a default measurement interval of 20 seconds.
During measurements, NMT setup functions (except for alarms) are not available and
appear ghosted in the NMT menu. To change NMT settings after startup, stop
measurements, change the settings and then restart measurements.
To stop NMT measurements
Click on
Stop NMT
.
Clear NMT References
To clear the stored supramaximal current and reference twitch, click on
Clear NMT
References
. After you clear NMT references, new references are automatically
recalculated at the next measurement.
NOTE:
To stop stimulation in an emergency disconnect the PodCom cable from the
monitor.
NOTE:
If the pod cannot establish reference values or if you wish to start over, clear the
stored references and repeat the startup procedure.
Reference values should generally be obtained during start-up, before you
administer muscle relaxants.
NOTE:
Only use this selection before patient is relaxed. If you clear NMT references
after the patient is already relaxed, the new references will be based on the existing level
of patient relaxation, which could cause misleading NMT measurements.
NOTE:
Changing the stimulation current or pulse width also clears stored references.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...