You can set TOF-count alarm limits and alarm recordings in the Alarm Limits table
(see chapter 5, Alarms).
Post-Tetanic Count (PTC)
In this mode, the pod sends stimuli at a 50 Hz rate for
5 seconds, pauses for 5 seconds and then sends single
pulses at a 1 Hz rate. This continues until no twitches are
detected or 20 stimuli are sent. The pod counts the number
of detected twitch responses and displays the post-tetanic
count in the NMT parameter box. The fewer responses
detected, the deeper the neuromuscular blockade.
Measurement intervals are not available in the PTC mode. If you want to return to
automatic measurement cycles after a PTC measurement, select the single or TOF
monitoring mode and restart measurements.
Taking NMT Measurements
To access the NMT Setup menu
1. Click on the
NMT
parameter box on the
Main Screen (if displayed).
or
1. Press the
Menu
fixed key.
2. Click on
Patient Setup
.
3. Click on
Parameters
.
4. Click on
NMT
.
NOTE:
There is no AutoSet-function for TOF-count alarm limits.
NOTE:
You can use PTC stimulation to measure deep neuromuscular blockade when
no twitch responses are detected in single or TOF mode.
If PTC measurements start showing a decrease of neuromuscular blockade,
change to Single or TOF mode as soon as possible in order to limit the number
of PTC measurements.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...