24-4
D
ELTA
/D
ELTA
XL/K
APPA
VF8
24 N
EUROMUSCULAR
T
RANSMISSION
(NMT)
MONITORING
Connections
1. Attach the distal (black) and proximal
(red) electrodes above the Ulnar
nerve as shown (3 to 5 cm. apart).
2. Connect the Trident pod with the
PodCom cable to the monitor’s
PodCom connector.
3. Apply the NMT hand adaptor.
4. Connect electrode and sensor cables
to the NMT cable (if applicable).
5. Connect the NMT cable to the
Trident pod (NMT connector).
1
NMT intermediate cable
2
Red (positive) electrode grabber
3
Black (negative) electrode grabber
4
NMT hand adapter
5
PodCom cable
NOTE:
Check that the thumb can move unconstrained before applying NMT sensor.
An EEG or BisX pod can connect to the monitor directly or via the NMT pod's
EEG connector. See chapter 11 for details about EEG monitoring and chapter
25 for details about BISx monitoring.
Use the provided hand adaptor straps to secure the hand adaptor.
1
Stimulus Indicator (LED)
2
NMT Cable connection port
3
To EEG pod or BISx pod
4
PodCom connection (to
monitor)
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...