![Dräger Infinity Delta Instructions For Use Manual Download Page 401](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779401.webp)
H
ARDWARE
S
ETUP
VF8
D
ELTA
/D
ELTA
XL/K
APPA
23-11
4. Connect one end of the Scio connecting cable to the X2 connector at the rear
of the module and the other end to the Delta/Delta XL/Kappa monitor or to
the IDS (see “Cable Connections” on page 23-9). If you are using a breakout
box, connect the end to the connector on the breakout box.
5. Connect the power supply to the external power supply connector at the rear
of the module.
6. Connect the power supply cord to a hospital outlet.
Warm-Up
Upon start-up, the Scio module passes through an initialization and warm-up period.
During this time, the etCO
2
* (in some models O
2
) and/or agent parameter boxes
display a question mark.
NOTE:
The exhaust port is a hose barb type connector.
CAUTION:
During warm-up, reported values might not be accurate. To achieve full
accuracy, a typical warm-up period of 7 minutes is recommended.
1
Bacterial filter
2
To sample gas recirculation sys-
tem (or gas scavenging system)
3
Sample gas return tubing
4
Exhaust port
5
External power supply connector
6
X2 connector
7
RS232C connector (X2 9 pin)
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...