Charging the Batteries
Before using the Delta/Delta XL monitor for the first time, charge the internal battery
for a maximum of 6.5 hours and the external battery for 3.5 hours.
When the monitor receives DC power, batteries recharge automatically. The optional
SLA Battery Charger can charge four fully depleted external batteries simultaneously
in approximately 3.5 hours. To start a fast-charge, insert the batteries into the slots of
the battery charger with the metal contacts facing down.
Power Sources (Kappa)
The Kappa monitor uses AC power (100-240 VAC). In case of a line outage or
disconnected cable, the monitor automatically switches to battery power to provide
continued patient monitoring without any loss of data. The base unit continues to
operate and alarm on battery power for approximately 20 minutes.
An internal battery can power the Kappa
monitor (but not the Kappa video display)
for up to approximately 20 minutes depending on the monitoring setup. It is intended
for short-term use only, for example, as a backup during power interruptions.
NOTE:
For the SC 7000/9000XL internal lead-acid battery, charge for a maximum of
4.5 hours the first time.
CAUTION:
The battery charger display is only accurate if the batteries are in normal
working condition.
WARNING: Read the sections “Safety Considerations” on
page i-6” and “Kappa Video Display” on page 1-11 of these
Instructions for Use before connecting the monitor to a power
source.
WARNING: Worn out or defective batteries can significantly
reduce this time.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...