22 F
I
O2 (F
RACTIONAL
I
NSPIRED
O2)
MONITORING
Menu Access
The FiO
2
menu displays the date and time of 1-point and 2-point calibrations (see
below). The menu item
Last O2 Cal
is informational only. It displays the date and
time of the last successful calibration, either 1-point or 2-point.
To open the FiO
2
menu
Click on the
FiO
2
parameter box on the Main screen
or
1. Press the
Menu
fixed key.
2. Click on
Patient Setup
.
3. Click on
Parameters
.
4. Click on
FiO2
.
Calibration
Every time a sensor is connected to the NeoMed pod, you must calibrate the monitor
to the sensor. The monitor does not display FiO
2
values until it is calibrated.
There are two types of calibration. 1-point calibration measures the oxygen in room
air, typically 21 %, and calibrates the monitor to that measurement. A 2-point
calibration uses two measurements, room air and 100 % oxygen, to calibrate the
monitor. A 2-point calibration provides more accurate FiO
2
monitoring because the
monitor is calibrated to two different measurements.
1-point calibration should be performed daily. A 2-point calibration should be
performed weekly. You should also calibrate the monitor under the following
circumstances:
Periodically, to verify the correct functioning of the sensor
Daily, if you are monitoring a patient’s FiO
2
on a daily basis
When you suspect that sensor characteristics have changed
When the accuracy of the monitor is in question
When there is a change in humidity or barometric pressure of the monitoring
site
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...