Infinity Docking Station (IDS)
The Infinity Docking Station (IDS) helps facilitate patient transport, allowing you to
remove the monitor from the bedside and dock it at another station while maintaining
patient and monitor connections. This feature, called Pick and Go, is explained in
further detail on page 3-5. With its companion DC power supply, the IDS provides
power and data connection, stores setup defaults, and connects your monitor to a
network.
Battery Power
The Delta/Delta XL monitor operates on an external, sealed lead-acid battery and an
internal lithium ion battery. The external battery, which can easily be replaced when
depleted, can power the monitor for approximately 50 minutes. If it runs low or you
remove it from a monitor that has been using battery power, the monitor automatically
switches to an internal battery, which can power the monitor for approximately 180
minutes (see page B-8).
NOTE:
The SC7000/9000XL monitor's internal lead-acid battery can power the
monitor for approximately 75 minutes.
1
MultiGas module
2
Power supply
3
Export protocol/alarm output
(nurse call)
1
, Scio module,
(X3)
4
Independent Surgical Display
(ISD) (X4)
5
Network
6
R50 series recorder (X13)
7
Video out / Remote display
(X5)
8
Potential equalization
9
MIB connectors
I
nfinity Docking Station
(Rear View
)
1
NOTE
: Alarm output (nurse call) requires the
alarm output cable with partnumber 5194928
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...