Purge Sensor Function
The etCO
2
/Respiratory Mechanics pod accommodates automatic or manual purging
of the flow sensor tubing. While the pod automatically purges the breathing circuit
periodically; manual purges are sometimes necessary.
Automatic Purging
The following table describes automatic purge characteristics.
Manual Purging
The
Paw
and
Vent
menu and
Purge Sensor
function allows manual purging between
the automatic purge cycles. See “Quick Reference Table: Paw and Vent Setup” on
Paw and Vent Setup Menus
On the Airway Pressure (
Paw
) and Flow/Volume (
Vent
) setup menus you can
configure the monitor to display respiratory mechanics parameters.
To access respiratory mechanics parameters setup menus:
Click on the parameter box labeled
Paw
on the main screen to open the
Airway Pressure setup menu or
Vent
to open the Flow/Volume setup menu.
Each of these setup menus has a second “page” that is displayed by clicking on the
more
arrow at the bottom of the screen.
NOTE:
Pressure and flow parameters are frozen and waveforms are not drawn during
purging.
Mode
Ventilator Circuit
Pressure
Interval
Adult
Adult/Pediatric
0 - 40 cmH2O
10 minutes
41 - 60 cmH2O
5 minutes
> 60 cmH2O
2 minutes
Pediatric
Not applicable
5 minutes
1
Neonatal
Not applicable
3 minutes
1
1
NOTE:
Sensor tubing lines are purged at required interval regardless
of circuit pressure.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...