Status Messages
Message
Condition
Suggested Action
etCO
2
<LL
etCO
2
> UL
Values are outside the current alarm
limits.
• Check the patient and treat, if
necessary.
• Change alarm limits.
• Check accessory and replace, if
necessary.
iCO
2
< LL
iCO
2
> UL
RRc < LL
RRc > UL
etCO2 Out of
Range (High)
The CO
2
value is out of range (high). Check the patient and treat, if
necessary.
RRc Out of
Range (High)
The RRc value is out of range (high).
etCO
2
Tubing
Disconnected
The FilterLine™ is unplugged.
Check sample line and reconnect.
etCO
2
Tubing
Leak
The sidestream tubing has a leak.
NOTE:
Occurs only during power up
or during an auto zero.
Replacethe tubing.
etCO
2
Tubing
Blocked
The FilterLine™ is obstructed.
• Disconnect and reconnect the
FilterLine™.
• Disconnect and replace with new
FilterLine™.
etCO
2
Zeroing
Autozero in progress
Informational message; no action
required.
etCO
2
Zero
Failed
Autozero unsuccessful.
• Retry to autozero.
• Return the pod to the hospital’s
technical personnel.
etCO
2
Zero
Accepted
Autozero completed successfully.
Informational message; no action
required
Calibrate etCO
2
pod
An initial calibration is requested after
the first 1,200 operating hours.
Afterwards, calibration is required
annually or at 4,000 operating hours,
whichever is first.
• Recalibrate pod.
• Return the pod to the hospital’s
technical personnel.
etCO
2
Calibrating
Calibration in progress.
Informational message; no action
required
etCO
2
Cal.
Accepted
Calibration process completed
successfully.
etCO
2
Cal.
Canceled
Calibration cancelled.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...