![Dräger Infinity Delta Instructions For Use Manual Download Page 341](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779341.webp)
ET
CO2 S
ETUP
VF8
D
ELTA
/D
ELTA
XL/K
APPA
19-11
etCO
2
Setup
The etCO
2
setup menus differ slightly depending on whether the monitor is connected
to an etCO
2
module or an etCO
2
pod. The procedure for opening the setup menu,
however, is the same for both devices.
Accessing Setup Menus
Click on the main screen etCO
2
parameter box to open the etCO
2
setup menu
or
1. Press the
Menu
fixed key on the front of the monitor. The Main menu
appears.
2. Click on
Patient Setup
. A list of available patient setup functions appears.
3. Click on
Parameters
in the second column. A list of available parameters
appears.
4. Click on
etCO2
. The etCO
2
setup menu appears.
Change in slope of
ascending limb. Possible
absence of an alveolar
plateau
Obstruction in Apparatus
• Partial obstruction in expiratory limb of
breathing circuit
• Foreign body in upper airway
• Partially kinked or occluded artificial
airway
• Herniated endotracheal/tracheostomy
tube cuff
• Bronchospasm
Elevated baseline, with
pronounced slope on
descending limb
Faulty Ventilator Circuit Valve
• Rebreathing (see above)
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...