19-6
D
ELTA
/D
ELTA
XL/K
APPA
VF8
19
ET
CO2 (E
ND
-T
IDAL
CO2)
MONITORING
Sampling Methods
Mainstream
For mainstream detection, the sensor is located within the patient’s airway or
breathing circuit, allowing you to monitor individual breathing cycles of the intubated
patient. This method is appropriate for neonatal as well as adult and pediatric patients.
Mainstream Monitoring Setup
1. Click on etCO
2
parameter box to access
etCO
2
setup menu.
2. Click on
Measurement Mode
.
3. Click on
Main
.
4. Select a mainstream airway adapter.
Make sure the windows are clean and
dry. Clean or replace the adapter if
necessary.
5. Align the marks on bottom of the adapter
with the sensor.
6. Snap firmly into place.
7. Insert the adapter in a vertical position
between the elbow and ventilator circuit “Y.
8. Position sensor cable away from patient.
9. Confirm proper connection on the etCO
2
setup menu.
NOTE:
During nebulization or suction for intubated patients, remove the sampling line
luer connector from the monitor to avoid moisture buildup and sampling line
occlusion.
Replace the sampling line according to hospital protocol or when a blockage is
indicated by the device. Excessive patient secretions or a build-up of liquids in
the airway tubing may occlude the sampling line, requiring more frequent
replacement.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...