Hardware
The transcutaneous monitoring system consists of a tpO
2
/CO
2
pod, a solid-state
sensor, and a calibration unit. The sensor contains two temperature sensors, a heating
element, and a combined O
2
and CO
2
sensing element. The sensor heats the patient’s
skin, increasing local blood perfusion. Oxygen and carbon dioxide diffuse through the
heated skin and the sensor measures their values on the skin’s surface. In neonates,
episodes of hypoxia and hyperoxia can therefore be quickly detected and avoided.
Sensor Warm-Up
The sensor’s heating element must warm the underlying tissue before obtaining
accurate tpO
2
/CO
2
values. The length of the warm-up period varies; it generally takes
10 to 15 minutes. Values displayed in the tpO
2
2/CO
2
parameter box during calibration
and the initial warm-up phase are not clinically useful because the reading is only
stable after the measurement site is warm and local hyperemization is complete.
tpO
2
/CO
2
alarms are disabled during warm-up. The start and end of the warm-up
period are marked on the trend display with a vertical line on the trend graph and an
event marker on the trend table.
WARNING: If the monitor reports a hardware failure, remove
the sensor immediately and do not reapply the sensor until
the problem has been corrected.
1
Calibration fixed key
2
Calibration chamber
3
Connection to tpO
2
/CO
2
electrode
4
Connection to monitor
tpO
2
/CO
2
pod
1
Heating element
2
Solid-state ceramic Severinghaus-
type CO
2
electrode
3
Platinum Clark-type O
2
electrode
4
Temperature sensors (2)
tpO
2
/CO
2
sensor
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...