17-10
D
ELTA
/D
ELTA
XL/K
APPA
VF8
17 P
ULSE
O
XIMETRY
(S
P
O2)
Status Messages
Monitor with Masimo Logo - Status Messages
SpO2 Desat Limit
This feature disables the SpO
2
desaturation alarm or sets a threshold at
which point a desaturation alarm of high
priority is triggered. For example, if the
lower alarm limit for SpO
2
is set at 95%
and the desaturation limit is set at –5, the
desaturation alarm will be triggered
when the patient’s SpO
2
value goes
below 90%.
NOTE
: This selection is available for all
patient categories and is protected by a
clinical password.
•
–5
,
–10
(default),
–15
,
–20
• OFF
SpO2 Alarm
Accesses SpO
2
alarms on the Alarm
Limits table (see page 5-6).
Not applicable
Message
Possible Cause
Suggested Action
SpO
2
> #
SpO
2
< #
The patient’s SpO
2
falls outside
the current upper or lower alarm
limits.
Observe the patient and treat, if
necessary.
PLS > #
PLS < #
The patient’s pulse rate falls
outside the current upper or
lower alarm limits.
SpO2 < Low limit
value - 20%
The patient’s SpO
2
falls below
the current lower alarm limit by
20% or more.
PLS Out of Range
(Low/High)
The pulse rate is outside the
measuring range of the monitor
SpO
2
Transparent
Too much light is reaching the
sensor’s light detector because
because the sensor is either
disconnected or too much
ambient light is present.
• Make sure the sensor is properly
attached to the patient’s finger.
• Remove the light source.
• Cover the sensor with opaque material.
• Make sure no ambient light can reach
the detector.
• Contact DrägerService.
Menu Item
Description
Settings
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...