Oxygenation and Ventilation Parameters
The monitor calculates the oxygenation and ventilation values using these parameters.
The monitor automatically calculates the following oxygenation and ventilation
parameters:
TPR
Total Pulmonary
Resistance
80 x PA M / CO
dynes x sec/cm
5
TVR
Total Vascular
Resistance
80 x ART M / CO
dynes x sec/cm
5
Label
Parameter Description
Derivation
Units
Hgb
Hemoglobin concentration
Monitored input
g/dl
iO2
Inspired Oxygen
Monitored input
%
PaCO2
Arterial CO2 Pressure
Input data
mmHg
PaO2
Arterial Oxygen Pressure
Input data
mmHg
PAUSE
Pause/Plateau Pressure
Monitored input
cmH2O
Pb
Barometric pressure
Monitored input
mmHg
PeCO2
Mixed Expired CO2 Pressure
Monitored input
mmHg
PEEP
Peak End Expiratory Pressure
Monitored input
cmH2O
PIP
Peak inspiratory Pressure
Monitored input
cmH2O
RRc, RRc*,
RRv
Respiratory Rate
Monitored input
L/m
SaO2,
SaO2*
Arterial Oxygen Saturation
Input data
%
SvO2
Venous Oxygen saturation
Monitored input
%
TVe
Expired Tidal Volume
Monitored input
ml/breath
Label
Parameter Description
Equations
Units
C(a-v)O
2
Arteriovenous Oxygen Difference
CaO
2
- CvO
2
ml/dl
CaO
2
Arterial Oxygen Content
.0134 x HGB x SaO2
ml/dl
C
DYN
Dynamic Compliance
C
DYN =
TVe/ (PIP - PEEP)
ml/cmH2O
Cs
Static Compliance
TVe/(Pause-PEEP)
ml/cmH2O
CvO
2
Venous Oxygen Content
.0134 x HGB x SvO2
ml/dl
DO
2
Oxygen Availability, Delivery, or
Transport
CaO
2
x CO x 10
mL/min.
DO
2
I
Oxygen Availability
(or Delivery)
Index
DO2 / BSA
mL/min/m
2
MV alv
Alveolar Minute Volume
(TVe- Vd aw) x RR
mL/min
Label Parameter
Description
Derivation
Units
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...